Literature DB >> 16886029

The efficacy of a two needle electrode technique in percutaneous radiofrequency rhizotomy:An investigational laboratory study in an animal model.

Richard Derby1, Chang-Hyung Lee.   

Abstract

BACKGROUND: Radiofrequency neurotomies are used to help reduce pain caused by structures innervated by the medial branch of the dorsal spinal nerve. The duration of effect may vary proportionally to the length of nerve coagulated. Techniques used to maximize the length of nerve within the radius of maximal heat include making multiple lesions, using larger needles, positioning the exposed needle parallel to the target nerve, and attempting more precise placement using 50 hertz test stimulation. A variation of the technique uses two needles that are simultaneously placed to lie parallel to one another and parallel to the probable area the target nerve is known to traverse. Heating both needles at the same time would be a faster method that theoretically might also include more tissue within the heating radius of the needle lengths.
OBJECTIVE: The purpose of this study was to confirm that two parallel radiofrequency lesions increased the volume of tissue included within the heating radius of the exposed needle tips, to determine optimal heating parameters, and to establish how far apart the needles can be placed and still achieve temperatures between the needles that will coagulate nerves. STUDY
DESIGN: Investigational laboratory study in an animal model.
METHODS: Ex-vivo porcine spinal tissue samples were prepared following a standardized protocol and two needle electrodes were connected to the Stryker rhizotomy device. A series of tissue samples were prepared with the needles placed 1, 1.5, 2, 3, 4, 5, 6, 7, 8, or 9 mm apart in respective samples. The needles were positioned parallel to the surface, and perpendicular to the tissue. In one series of samples, two needle electrodes were heated sequentially for each needle placement. Temperature parameters ranged from 60 to 100 degrees C and the time duration ranged from 60 to 90 seconds for each electrode. After each heating, all tissue was dissected and the size of the coagulated area measured. Temperatures at the midpoint between the two needle electrodes were monitored and measured with a Radionics needle sensor. Using the same study protocol and evaluation, a second set of samples was studied but both needle electrodes were heated simultaneously.
RESULTS: Maximal lesion size was observed when the temperature of the electrode was higher than 70-80 degrees C and maintained for longer than 70 seconds. When the needles were heated sequentially, the needles could be placed no further than 1.5 to 2 mm apart before areas of incomplete coagulation were visualized at the midpoint between the needles. When both needles were simultaneously heated the needles could be placed 4 mm apart before incomplete coagulation was visualized. With simultaneous heating, the highest peak temperature of the midpoint between two needles (6 mm apart) exceeded 66 degrees C, compared to a temperature of 40 degrees C achieved with sequential heating.
CONCLUSIONS: This research confirms the efficacy of utilizing two needle electrodes during lumbar facet rhizotomy in an experimental model. Heating the two electrodes sequentially appears to coagulate a wider area and thus would be more likely to include a longer length of the target nerve and thus potentially could achieve better results in less time.

Entities:  

Mesh:

Year:  2006        PMID: 16886029

Source DB:  PubMed          Journal:  Pain Physician        ISSN: 1533-3159            Impact factor:   4.965


  6 in total

1.  Sequential bipolar radiofrequency lumbar sympathectomy in Raynaud's disease -A case report-.

Authors:  Sang-Soo Kang; Keun-Man Shin; Sang-Moon Jung; Jun-Hee Park; Seong-Jun Hong
Journal:  Korean J Anesthesiol       Date:  2010-10-21

2.  Morphologic Analysis of Water-Cooled Bipolar Radiofrequency lesions on Egg White in Vitro.

Authors:  Sang Soo Kang; Jung Chan Park; Young Jun Yoon; Keun Man Shin
Journal:  Korean J Pain       Date:  2012-06-28

3.  Endoscopically guided foraminal and dorsal rhizotomy for chronic axial back pain based on cadaver and endoscopically visualized anatomic study.

Authors:  Anthony Yeung; Satishchandra Gore
Journal:  Int J Spine Surg       Date:  2014-12-01

4.  Clinical Identification of the Vertebral Level at Which the Lumbar Sympathetic Ganglia Aggregate.

Authors:  Ji Won An; Jae Chul Koh; Jong Min Sun; Ju Yeon Park; Jong Bum Choi; Myung Ju Shin; Youn Woo Lee
Journal:  Korean J Pain       Date:  2016-04-01

5.  Fluoroscopy-guided Bipolar Radiofrequency Thermocoagulation Treatment for Discogenic Low Back Pain.

Authors:  Li Zhang; Xin-Li Ding; Xu-Li Zhao; Jun-Nan Wang; Yan-Ping Li; Ming Tian
Journal:  Chin Med J (Engl)       Date:  2016-10-05       Impact factor: 2.628

6.  Analysis of radiofrequency lesions in egg whites in vitro produced by application of the Tew electrode for different temperatures and times.

Authors:  Young Suk Kwon; So Young Lim; Jong Ho Kim; Ji Su Jang; Chul Ho Kim; Kwang Jun Kwon; Jun Heum Yon
Journal:  Pain Res Manag       Date:  2015-09-10       Impact factor: 3.037

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.