Literature DB >> 16880163

3D non-linear analysis of the acetabular construct following impaction grafting.

A T M Phillips1, P Pankaj, C R Howie, A S Usmani, A H R W Simpson.   

Abstract

The study investigates the short-term behaviour of the acetabular construct following revision hip arthroplasty, carried out using the Slooff-Ling impaction grafting technique; using 3D finite element analyses. An elasto-plastic material model is used to describe the constitutive behaviour of morsellised cortico-cancellous bone (MCB) graft, since it has been shown that MCB undergoes significant plastic deformation under normal physiological loads. Based on previous experimental studies carried out by the authors and others, MCB is modelled using non-linear elasticity and Drucker Prager Cap (DPC) plasticity. Loading associated with walking, sitting down, and standing up is applied to the acetabular cup through a femoral head using smooth sliding surfaces. The analyses yield distinctive patterns of migration and rotation due to different activities. These are found to be similar to those observed in the clinical setting.

Entities:  

Mesh:

Year:  2006        PMID: 16880163     DOI: 10.1080/10255840600732226

Source DB:  PubMed          Journal:  Comput Methods Biomech Biomed Engin        ISSN: 1025-5842            Impact factor:   1.763


  1 in total

1.  [Biomechanical analysis and effectiveness evaluation of zone ++ reconstruction of hemipelvis with rod-screw prosthesis].

Authors:  Jingyi Dang; Zhao Zhang; Zhenzhou Mi; Debin Cheng; Jun Fu; Dong Liu; Hongbin Fan
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2022-04-15
  1 in total