Toxic vessel reaction to an absorbable polymer-based paclitaxel-eluting stent in pig coronary arteries.

Refat Jabara1, Nicolas Chronos, Fernando Tondato, Damian Conway, Warner Molema, Kenneth Park, Tom Mabin, Spencer King, Keith Robinson.   

Abstract

OBJECTIVES: The goal of this study was to evaluate a new drug-eluting stent (DES) comprising a bioabsorbable polymer eluting a moderate dose of paclitaxel in a clinically relevant animal model.
BACKGROUND: Although DES limit restenosis, adverse vascular pathologies and toxicities continue to be of major concern. Optimization of DES components, especially completely absorbable polymers, may reduce these toxicities.
METHODS: Bare-metal (BM), absorbable polymer coating only (POLY), and polymer-based paclitaxel-eluting (PACL) stents were implanted in porcine coronary arteries using intravascular ultrasound (IVUS) to optimize stent apposition. The dose density of paclitaxel was 0.30-0.35 mcg/mm2, with in vitro elution studies demonstrating a gradual elution over 6-8 weeks. The animals were terminated at 1 week, 1 month and 3 months. Histopathologic and histomorphometric analyses were perform.
RESULTS: The arteries with PACL showed extensive smooth muscle cell necrosis at 1 week and poor apposition of stent struts at 1 month (malapposition measured as gap width between strut and internal elastic lamina), with greater gap width compared to the BM and POLY groups (0.22 mm +/- 0.02 vs. 0.03 mm +/- 0.02 and 0.02 mm +/- 0.01, respectively; p < 0.001). At 3 months, the PACL group showed rebound neointimal thickness and histological percent stenosis compared to the BM group (0.48 mm +/- 0.14 vs. 0.07 mm +/- 0.02, respectively; p < 0.001 and 59% +/- 11 vs. 17% +/- 2, respectively; p < 0.001).
CONCLUSIONS: Despite in vitro data showing slow, sustained release of paclitaxel from a bioabsorbable polymer, the porcine coronary artery model demonstrated a sequence of medial necrosis, stent malapposition and late neointimal thickening. Since the therapeutic window for paclitaxel may be narrower than currently inferred, thorough preclinical testing coupled with the polymer development process for stents eluting paclitaxel is needed.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16877788

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  2 in total

1.  In vivo Evaluation of Cenderitide-Eluting Stent (CES) II.

Authors:  Yingying Huang; Xu Wen Ng; Soon Ghim Lim; Horng Haur Chen; John C Burnett; Yin Chiang Freddy Boey; Subbu S Venkatraman
Journal:  Ann Biomed Eng       Date:  2015-07-16       Impact factor: 3.934

2.  Drug-eluting stents in preclinical studies: updated consensus recommendations for preclinical evaluation.

Authors:  Robert S Schwartz; Elazer Edelman; Renu Virmani; Andrew Carter; Juan F Granada; Greg L Kaluza; Nicolas A F Chronos; Keith A Robinson; Ron Waksman; Judah Weinberger; Gregory J Wilson; Robert L Wilensky
Journal:  Circ Cardiovasc Interv       Date:  2008-10       Impact factor: 6.546

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.