Phacodynamics: an aspiration flow vs vacuum comparison.

Wesley Adams1, Jason Brinton, Michael Floyd, Randall J Olson.   

Abstract

PURPOSE: To compare actual flow and the vacuum necessary to generate that flow for the Advanced Medical Optics Sovereign (Advanced Medical Optics, Santa Ana, California, USA) and the Alcon Legacy (Alcon, Fort Worth, Texas, USA) phacoemulsification machines.
DESIGN: In vitro laboratory study.
METHODS: Flow was collected and machine indicated vacuum noted from 12 ml/min to 40 ml/min in 2 ml/min steps (20-gauge). Nineteen and 20-gauge tips were also compared. Machine indicated vacuum was converted to actual tip vacuum.
RESULTS: Legacy flow (102.4 +/- 3.8% of indicated) was significantly more than Sovereign (96.5 +/- 3.9%; P < .0001). 20-gauge flow was less than 19-gauge flow (96.1% vs 98.6% of indicated, P < .0001). Sovereign had less unoccluded vacuum (70.3 +/- 3.2 mm Hg) than Legacy (89.4 +/- 1.2 mm Hg) at 36 ml/min of flow (P < .0001 for 26 to 36 ml/min of flow).
CONCLUSION: There is unoccluded vacuum at the phaco tip of these systems at 14 ml/min, of flow and at 40 ml/min, this was 112.0 mm Hg for the Legacy.

Entities:  

Mesh:

Year:  2006        PMID: 16876517     DOI: 10.1016/j.ajo.2006.02.034

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  2 in total

1.  Measurement of Phacoemulsification Vacuum Pressure in the Oertli CataRhex3.

Authors:  John S Peterson; Caitlynn Cooper; Emilie L Ungricht; Christina Mamalis; William Barlow; Brian Zaugg; Ashlie A Bernhisel; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2022-06-01

2.  Real-time intraocular pressure measurement during phacoemulsification in dogs ex vivo.

Authors:  Seonmi Kang; Sangwan Park; Hyunwoo Noh; Jiyoon Kwak; Kangmoon Seo
Journal:  J Vet Med Sci       Date:  2015-02-16       Impact factor: 1.267

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.