Jim X Shen1, Cristina I Tama, Roger N Hayes.
Abstract
Automated micro-SPE tips were successfully utilized for the determination of posaconazole in rat plasma. The bioanalytical method using micro-SPE tips was successfully qualified for routine quantitation of posaconazole over the concentration range of 10.0-10,000 ng/mL in rat EDTA plasma. Inter-assay precision, based on percent relative deviation for n=18 replicate quality controls, was < or =5.7%. Inter-assay accuracy based on n=18 replicate quality controls was +/-7.7%. Complete solid phase extraction using micro-SPE tips was demonstrated on a Tomtec liquid handler where >95% recovery for posaconazole was obtained. The micro-SPE tips had sufficient capacity to extract at least 100 microL plasma fortified with 10 microg/mL of posaconazole and the analyte could be efficiently eluted with as little as 60 microL of methanol. Of particular note is the unique ability of these micro-SPE tips to perform exhaustive solid phase extraction more commonly performed when using liquid/liquid extraction.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16872919 DOI: 10.1016/j.jchromb.2006.06.039
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205