Literature DB >> 16872086

Wall correction factors, Pwall, for parallel-plate ionization chambers.

Lesley A Buckley1, D W O Rogers.   

Abstract

The EGSnrc Monte Carlo user-code CSnrc is used to calculate wall correction factors, Pwall,, for parallel-plate ionization chambers in photon and electron beams. A set of Pwall values, computed at the reference depth in water, is presented for several commonly used parallel-plate chambers. These values differ from the standard assumption of unity used by dosimetry protocols by up to 1.7% for clinical electron beams. Calculations also show that Pwall is strongly dependent on the depth of measurement and can vary by as much as 6% for a 6 MeV beam in moving from a depth of dref to a depth of R50. In photon beams, where there is limited information available regarding Pwall for parallel-plate chambers, CSnrc calculations show Pwall values of up to 2.4% at the reference depth over a range of photon energies. The Pwall values for photon beams are in good agreement with previous estimates of the wall correction but have much lower statistical uncertainties and cover a wider range of photon beam energies.

Entities:  

Mesh:

Year:  2006        PMID: 16872086     DOI: 10.1118/1.2199988

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Monte Carlo calculations of the replacement correction factor, P(repl), for cylindrical chamber cavities in clinical photon and electron beams.

Authors:  Fujio Araki
Journal:  Radiol Phys Technol       Date:  2012-04-22

2.  Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector.

Authors:  Frédéric Lacroix; Mathieu Guillot; Malcolm McEwen; Claudiu Cojocaru; Luc Gingras; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

  2 in total