Pacing and sensing disturbances in patients with DDD pacemakers in the early period after implantation.

Jerzy Krzysztof Wranicz1, Michał Chudzik, Iwona Cygankiewicz, Artur Klimczak, Krzysztof Kaczmarek, Marek Maciejewski, Jan Henryk Goch.   

Abstract

OBJECTIVES: We sought to determine the usefulness of ambulatory 24-hour Holter monitoring in detecting asymptomatic pacemaker (PM) malfunction episodes in patients with dual-chamber pacemakers whose pacing and sensing parameters were proper, as seen in routine post-implantation follow-ups. This aspect has not been widely discussed so far. METHODS AND
RESULTS: Ambulatory 24-hour Holter recordings [HM] were performed in 100 patients with DDD pacemakers one day after the implantation. Only asymptomatic patients with proper pacing and sensing parameters (assessed on PM telemetry on the first day post-implantation) were enrolled in the study. The following parameters were assessed: failure to pace, failure to sense (both oversensing and undersensing episodes) as well as the percentage of all PM disturbances. Despite proper sensing and pacing parameters, HM revealed PM disturbances in 23% of the patients.Atrial undersensing episodes were found in 12 patients and failure to capture in I patient. T wave oversensing was the most common ventricular channel disorder (9 patients). Malfunction episodes occurred sporadically, leading to pauses of up to 1.6 s or temporary bradycardia, which were, nevertheless, not accompanied by clinical symptoms. No ventricular pacing disturbances were found.
CONCLUSION: Asymptomatic pacemaker dysfunction may be observed in nearly 25% of patients with proper DDD parameters after implantation.Thus, ambulatory HM during the early post-implantation period may be a useful tool to detect the need to reprogramme PM parameters.

Entities:  

Mesh:

Year:  2006        PMID: 16869449     DOI: 10.2143/AC.61.3.2014830

Source DB:  PubMed          Journal:  Acta Cardiol        ISSN: 0001-5385            Impact factor:   1.718


  1 in total

1.  [Atrial sensing with fixed and floating electrodes at identical activities].

Authors:  D Hartung; L Thiele; W M Hartung
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-04-18
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.