Repeatability of a computerized muscle tonometer and the effect of tissue thickness on the estimation of muscle tone.

Jari Ylinen1, Iiro Teittinen, Ville Kainulainen, Hannu Kautiainen, Kari Vehmaskoski, Arja Häkkinen.   

Abstract

The aim of the present study was to assess the intrarater repeatability of the computerized muscle tonometer (CMT), which is a computer-driven device designed to measure and register the tone of connective tissues. The second aim was to evaluate the effect of muscle thickness on results and repeatability as well as sensitivity, which was done by comparison to a mechanical hand-held tonometer (HHT). Participants were ten healthy male volunteers. The thickness of soft tissues was measured with ultrasound and the CMT was used to evaluate the muscle tone of thigh muscles 10, 15 and 20 cm and the HHT 15 cm above the patella on both sides. Great individual variability in muscle thickness and tone was noted. Muscle thickness increased proximally on the thigh and thus the mean distance travelled by the indenter of the CMT also increased from distal to proximal from 22 to 26 mm and the work done increased from 116 to 161 mJ. Intraclass correlation (ICC) for travel and work ranged from 0.85 to 0.92 and 0.94 to 0.98, respectively. Repeatability coefficients increased somewhat at proximal measurement sites. ICC for the HHT was 0.78. The CMT also showed a considerably higher range of results, indicating greater resolution power. Despite the twofold pressure used with the HHT the distance measured was on average 60% lower compared to the CMT. In conclusion, the repeatability of the CMT was found to be good. Muscle thickness greatly affected the results and should therefore be taken into consideration when measuring muscle tone. Work was a more reliable unit of measurement than travel, which has been commonly used to describe muscle tone.

Entities:  

Mesh:

Year:  2006        PMID: 16868346     DOI: 10.1088/0967-3334/27/9/003

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

1.  Effect of gender, muscle type and skinfold thickness on myometric parameters in young people.

Authors:  Joanna Mencel; Anna Jaskólska; Jaroslaw Marusiak; Katarzyna Kisiel-Sajewicz; Magdalena Siemiatycka; Lukasz Kaminski; Artur Jaskólski
Journal:  PeerJ       Date:  2021-11-10       Impact factor: 2.984

2.  Reduced Neck Muscle Strength and Altered Muscle Mechanical Properties in Cervical Dystonia Following Botulinum Neurotoxin Injections: A Prospective Study.

Authors:  Sirpa Mustalampi; Jari Ylinen; Katariina Korniloff; Adam Weir; Arja Häkkinen
Journal:  J Mov Disord       Date:  2016-01-25

3.  Exploring the Pathological Role of Collagen in Paravertebral Muscle in the Progression of Idiopathic Scoliosis.

Authors:  Haidong Peng; Feng Jin; Depeng Meng; Jun Li; Shuhan Yu; Shen Zhang; Guigang Zeng
Journal:  Biomed Res Int       Date:  2020-08-03       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.