Comparison of intraarterial MR angiography at 3.0 T with X-ray digital subtraction angiography for detection of renal artery stenosis in swine.

Thomas K Rhee1, Jonathan K Park, Ty A Cashen, Wanyong Shin, Brian E Schirf, James A Gehl, Andrew C Larson, James C Carr, Debiao Li, Timothy J Carroll, Reed A Omary.   

Abstract

PURPOSE: To compare the accuracy of catheter-directed intraarterial (IA) magnetic resonance (MR) angiography at 3.0 T with that of x-ray digital subtraction angiography (DSA) for the measurement of renal artery stenosis (RAS) in swine.
MATERIALS AND METHODS: Unilateral hemodynamically significant RAS (>50%) was induced surgically in six pigs with use of reverse cable ties. One to two weeks after surgery, each pig underwent x-ray DSA and MR angiography before and after percutaneous transluminal balloon angioplasty (PTA). X-ray DSA was performed before and after PTA of RAS by injection of iodinated contrast agent through a 5-F multiple-side hole angiographic catheter placed in the abdominal aorta under fluoroscopic guidance. MR angiography of RAS was performed before and after PTA of RAS on a 3.0-T clinical MR imager with use of gadolinium-based contrast agent. MR angiography and DSA images were analyzed with the full width at half maximum method. Percent stenosis measurements between x-ray DSA and MR angiography were compared with a paired t test and were correlated with linear regression and Bland Altman analysis (alpha = 0.05).
RESULTS: Six cases of RAS were induced and imaged successfully with DSA and MR angiography techniques before and after PTA. On x-ray DSA, median stenoses was 64% (95% CI 57%-80%) before PTA and 20% (95% CI 5%-32%) after PTA. Corresponding MR angiography median stenosis measurement was 69% (95% CI 58%-80%) before PTA and 26% (95% CI 16%-36%) after PTA. A paired t test comparison did not show a difference between DSA and MR angiography (P = .16). RAS measurements on MR angiography correlated closely (P < .01) with DSA measurements (r(2) = 0.92).
CONCLUSION: In swine, the accuracy of catheter-directed IA MR angiography with use of a clinical 3.0-T MR imaging unit for the measurement of RAS was similar to that of conventional x-ray DSA.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16868166     DOI: 10.1097/01.RVI.0000228469.10687.2C

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464