Radiofrequency ablation (RFA): development of a flow model for bovine livers for extensive bench testing.

Andreas Lubienski1, Rudi G Bitsch, Katrin Lubienski, Guenter Kauffmann, Markus Duex.   

Abstract

PURPOSE: To develop a flow model for bovine livers for extensive bench testing of technical improvements or procedure-related developments of radiofrequency ablation excluding animal experiments.
METHODS: The perfusion of bovine livers directly from the slaughterhouse was simulated in a liver perfusion tank developed for the experimental work. The liver perfusion medium used was a Tyrode solution prepared in accordance with physiologic criteria (as for liver transplants) which was oxygenated by an oxygenator and heated to 36.5 degrees C. Portal vein circulation was regulated via a flow- and pressure-controlled pump and arterial circulation using a dialysis machine. Flow rate and pressure were adjusted as for the physiology of a human liver converted to bovine liver conditions. The fluid discharged from the liver was returned into the perfusion system through the vena cava. Extendable precision swivel arms with the radiofrequency probe attached were mounted on the liver perfusion tank. RFA was conducted with the RF3000 generator and a 2 cm LeVeen needle (Boston Scientific, Ratingen, Germany) in a three-dimensional grid for precise localization of the generated thermolesions.
RESULTS: Four bovine livers weighing 8.4 +/- 0.4 kg each were prepared, connected to the perfusion system, and consecutively perfused for the experiments. Mean arterial flow was 569 +/- 43 ml/min, arterial pressure 120 mmHg, portovenous flow 1440 +/- 305 ml/min, and portal pressure 10 mmHg. Macroscopic evaluation after the experiments revealed no thrombi within the hepatic vessels. A total of 136 RF thermolesions were generated with an average number of 34 per liver. Mean RF duration was 2:59 +/- 2:01 min:sec with an average baseline impedance of 28.2 +/- 3.4 ohms. The mean diameter of the thermolesions along the puncture channel was 22.98 +/- 4.34 mm and perpendicular to the channel was 23.27 +/- 4.82 mm.
CONCLUSION: Extracorporeal perfusion of bovine livers with consecutive standardized RF ablation was feasible. The bovine liver flow model seems to allow extensive, standardized evaluation of technical or procedure-related developments of RF systems.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16845558     DOI: 10.1007/s00270-004-0338-4

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  4 in total

1.  Fast Estimation of the Vascular Cooling in RFA Based on Numerical Simulation.

Authors:  T Kröger; T Pätz; I Altrogge; A Schenk; K S Lehmann; B B Frericks; J-P Ritz; H-O Peitgen; T Preusser
Journal:  Open Biomed Eng J       Date:  2010-02-04

2.  Electrochemical treatment: An investigation of dose-response relationships using an isolated liver perfusion model.

Authors:  Ralf Czymek; Dorothea Dinter; Stephan Löffler; Maximilian Gebhard; Tilman Laubert; Andreas Lubienski; Hans-Peter Bruch; Andreas Schmidt
Journal:  Saudi J Gastroenterol       Date:  2011 Sep-Oct       Impact factor: 2.485

3.  Creation of an ex-vivo bovine kidney flow model for testing embolic agents: work in progress.

Authors:  Luis Garza; Ryan Bitar; Barrett O'Donnell; Matthew Parker; Carlos Ortiz; Charles Hyman; John Walker; Ho-Young Song; Jorge Lopera
Journal:  CVIR Endovasc       Date:  2021-02-03

4.  Evaluation of the in vivo efficiency and safety of hepatic radiofrequency ablation using a 15-G Octopus® in pig liver.

Authors:  Eun Sun Lee; Jeong Min Lee; Kyung Won Kim; In Joon Lee; Joon Koo Han; Byung Ihn Choi
Journal:  Korean J Radiol       Date:  2013-02-22       Impact factor: 3.500

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.