| Literature DB >> 16841279 |
Jacob Bornstein1, Adam Geva, Ido Solt, Vladimir Fait, Alex Schoenfeld, Hadar Kessary Shoham, Jack Sobel.
Abstract
This article describes the evaluation of the diagnostic efficacy of AL-SENSE panty-liner in detecting premature rupture of membranes (PROM). One hundred and three women attending the labor and delivery ward were enrolled into three groups: women presenting with a vaginal leak of fluid who had not yet been examined; women with overt PROM, and women with no fluid leak. The result of the AL-SENSE strip test was compared with the clinical diagnosis, which was based on direct visualization of the posterior vaginal fornix and cervix, crystallization, and nitrazine tests. AL-SENSE panty-liner test had a sensitivity of 100% and a specificity of 75% in detecting PROM, with an overall agreement of 82.35% between the AL-SENSE test result and the clinical diagnosis. AL-SENSE may be used as a reliable test to rule out PROM and as an effective device to diagnose PROM and differentiate it from urine leak and vulvovaginal candidiasis.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16841279 DOI: 10.1055/s-2006-947159
Source DB: PubMed Journal: Am J Perinatol ISSN: 0735-1631 Impact factor: 1.862