Literature DB >> 16837205

Gamma-ray imaging with a Si/CsI(Tl) Compton detector.

A S Hoover1, J P Sullivan, B Baird, S P Brumby, R M Kippen, C W McCluskey, M W Rawool-Sullivan, E B Sorensen.   

Abstract

We present results from Compton imaging of gamma-ray sources using an instrument constructed from thin silicon scattering detectors and CsI(Tl) absorbing detectors. We have successfully imaged single and double point sources for several common radioactive isotopes ((137)Cs, (60)Co, (22)Na, (54)Mn). The measured angular resolution is 11.6( composite function) FWHM at 662keV. In parallel with the hardware effort, a GEANT4-based simulation code was developed. Comparisons between real and simulated data are discussed.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16837205     DOI: 10.1016/j.apradiso.2006.05.015

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

Review 1.  Compton imaging for medical applications.

Authors:  Hideaki Tashima; Taiga Yamaya