Literature DB >> 16824282

Most commercial insulin assays fail to detect recombinant insulin analogues.

A H Heald1, B Bhattacharya, H Cooper, A Ullah, A McCulloch, S Smellie, G Wark.   

Abstract

Insulin assays are utilized in various clinical scenarios, including the assessment of insulin therapy compliance or of suspected insulin overdose. In an interpretative exercise carried out by UK National External Quality Assessment Service (NEQAS), serum sent to the participating laboratories was spiked with 30 pmol/L of the short-acting insulin analogue Human Actrapid. Only two out of 24 participant laboratories had sufficient assay cross-reactivity with Actrapid to interpret the results as suggestive of insulin administration. The development of specific insulin assays has led to deterioration in the ability to detect non-compliance or overdose with recombinant insulin treatment. Clinicians should be aware of this significant limitation, which could lead to misdiagnosis.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16824282     DOI: 10.1258/000456306777695690

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  10 in total

1.  Disorders of glucose metabolism: post mortem analyses in forensic cases--part II.

Authors:  Frank Musshoff; Cornelius Hess; Burkhard Madea
Journal:  Int J Legal Med       Date:  2010-10-07       Impact factor: 2.686

Review 2.  Clinical and Laboratory Aspects of Insulin Autoantibody-Mediated Glycaemic Dysregulation and Hyperinsulinaemic Hypoglycaemia: Insulin Autoimmune Syndrome and Exogenous Insulin Antibody Syndrome.

Authors:  Tony Huynh
Journal:  Clin Biochem Rev       Date:  2020-12

3.  Analysis of insulin in human breast milk in mothers with type 1 and type 2 diabetes mellitus.

Authors:  T J Whitmore; N J Trengove; D F Graham; P E Hartmann
Journal:  Int J Endocrinol       Date:  2012-03-05       Impact factor: 3.257

4.  Importance of insulin immunoassays in the diagnosis of factitious hypoglycemia.

Authors:  Özlem Nalbantoğlu Elmas; Korcan Demir; Nusret Soylu; Nilüfer Çelik; Behzat Özkan
Journal:  J Clin Res Pediatr Endocrinol       Date:  2014-12

5.  Factitious Hypoglycemia Caused by a Unique Pattern of Drug Use: A Case Report.

Authors:  Mojtaba Akbari; Akbar Soltani; Mohammad Reza Mohajeri-Tehrani; Hamidreza Aghaei Meybodi; Imaneh Sadat Mousavi Fakhr; Sayed Mahmoud Sajjadi-Jazi
Journal:  Int J Endocrinol Metab       Date:  2017-12-20

6.  Clinical Utility and Cross-Reactivity of Insulin and C-Peptide Assays by the Lumipulse G1200 System.

Authors:  Jongwon Oh; Jae Hyeon Kim; Hyung Doo Park
Journal:  Ann Lab Med       Date:  2018-11       Impact factor: 3.464

7.  Explaining unexplained hypoglycemia: How LC-MS/MS can help.

Authors:  M T Ackermans; J Hopman; A C Heijboer; S E Siegelaar
Journal:  Pract Lab Med       Date:  2022-07-12

8.  Insulin as a Tool in Factitious Dysglycemia.

Authors:  Samih A Odhaib; Qussay N Almaliki; Abbas A Mansour
Journal:  Cureus       Date:  2021-04-22

9.  Expression of the Biologically Active Insulin Analog SCI-57 in Nicotiana Benthamiana.

Authors:  Adriana Muñoz-Talavera; Miguel Ángel Gómez-Lim; Luis A Salazar-Olivo; Jörg Reinders; Katharina Lim; Abraham Escobedo-Moratilla; Alberto Cristian López-Calleja; María Cristina Islas-Carbajal; Ana Rosa Rincón-Sánchez
Journal:  Front Pharmacol       Date:  2019-11-14       Impact factor: 5.810

Review 10.  The Other Face of Insulin-Overdose and Its Effects.

Authors:  Szymon Rzepczyk; Klaudia Dolińska-Kaczmarek; Aleksandra Uruska; Czesław Żaba
Journal:  Toxics       Date:  2022-03-03
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.