Literature DB >> 16816760

Porous coated motion cervical disc replacement: a biomechanical, histomorphometric, and biologic wear analysis in a caprine model.

Nianbin Hu1, Bryan W Cunningham, Paul C McAfee, Seok Woo Kim, John C Sefter, Andrew Cappuccino, Luiz Pimenta.   

Abstract

STUDY
DESIGN: The biomechanical, histopathologic, and histomorphometric characteristics of cervical disc replacement were assessed in a caprine animal model.
OBJECTIVE: To investigate the biomechanical, porous ingrowth, and histopathologic characteristics of the Porous Coated Motion (PCM) Cervical Disc replacement (Cervitech, Inc., Rockaway, NJ). SUMMARY OF BACKGROUND DATA: As an alternative to anterior cervical interbody arthrodesis, an artificial cervical disc serves to replace the symptomatic degenerated disc, restore the functional biomechanical properties of the motion segment, and preserve neurologic function.
METHODS: There were 12 mature Nubian goats divided into 2 groups based on postoperative survival periods of 6 (n = 6) and 12 months (n = 6). Using an anterior surgical approach, a complete discectomy was performed at the C3-C4, followed by implantation of the PCM device. Functional outcomes of the disc prosthesis were based on computerized tomography (CT), multidirectional flexibility testing, undecalcified histology, histomorphometric, and immunocytochemical analyses.
RESULTS: There was no evidence of prosthesis loosening, or neurologic or vascular complications. CT showed the ability to image and assess the cervical spinal canal for the presence of compressive pathology in the area of the CoCrMo prosthesis. Multidirectional flexibility testing under axial rotation and lateral bending indicated no differences in the full range of intervertebral motion between the disc prosthesis and nonoperative controls (P > 0.05). Based on immunohistochemical and histologic analysis, there was no evidence of particulate debris, cytokines, or cellular apoptosis within the local or systemic tissues. Moreover, review of the spinal cord at the operative levels indicated no evidence of cord lesions, inflammatory reaction, wear particles, or significant pathologic changes in any treatment. Histomorphometric analysis at the metal-bone interface indicated the mean trabecular ingrowth of 40.5% +/- 24.4% and 58.65% +/- 28.04% for the 6 and 12-month treatments, respectively.
CONCLUSION: To our knowledge, this serves as the first in vivo time-course study investigating the use of the PCM device for cervical arthroplasty. All 12 animals undergoing cervical disc replacement had no evidence of implant loosening, subluxation, or inflammatory reactions. PCM cervical arthroplasty permits unobstructed visualization of the spinal canal based on CT imaging. Segmental intervertebral motion was preserved under axial rotation and lateral bending loading conditions, while at the same time permitting porous osseointegration at the prosthesis-bone interface. Based on histopathologic review of all local and systemic tissues, there was no evidence of particulate wear debris, cytokines, cellular apoptosis, or significant pathologic changes in any treatment.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16816760     DOI: 10.1097/01.brs.0000224537.79234.21

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  5 in total

Review 1.  Cervical and lumbar spinal arthroplasty: clinical review.

Authors:  T D Uschold; D Fusco; R Germain; L M Tumialan; S W Chang
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-27       Impact factor: 3.825

2.  Parathyroid hormone enhances gap healing and osseointegration in orthopedic porous coated titanium implants: a correlative micro-computed tomographic, histomorphometric and biomechanical analysis.

Authors:  Xinlin Gao; Yang Meng; Dingjun Hao; Hao Liu
Journal:  BMC Musculoskelet Disord       Date:  2022-01-03       Impact factor: 2.362

3.  A systematic review of cervical artificial disc replacement wear characteristics and durability.

Authors:  Ronald Lehman; Adam J Bevevino; Devon D Brewer; Andrea C Skelly; Paul A Anderson
Journal:  Evid Based Spine Care J       Date:  2012-02

4.  A comparison of the shock-absorbing properties of cervical disc prosthesis bearing materials.

Authors:  Michael C Dahl; Stephen Jacobsen; Newton Metcalf; Rick Sasso; Randal P Ching
Journal:  SAS J       Date:  2011-06-01

5.  In-vivo study of osseointegration in Prestige LP cervical disc prosthesis.

Authors:  Jigang Lou; Beiyu Wang; Tingkui Wu; Wenjie Wu; Huibo Li; Ziyang Liu; Hao Liu
Journal:  BMC Musculoskelet Disord       Date:  2018-02-07       Impact factor: 2.362

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.