J J Berzas Nevado1, M J Villaseñor Llerena, C Guiberteau Cabanillas, V Rodríguez Robledo.
Abstract
Capillary gas chromatography with mass spectrometry detection in SIM mode (GC-MS-SIM) has been used for the analysis of citalopram (CIT), fluoxetine (FLX), and all of their metabolites in urine samples. The instrumental parameters affecting GC separation and MS-SIM detection were investigated. A validation procedure was performed on urine matrix and a simultaneous robustness/ruggedness evaluation is also presented in this paper. An optimized solid-phase extraction (SPE) has been applied, reaching in this way to limits of detection (LODs) between 0.7 ng L(-1) (CIT) and 33.6 microg L(-1) (CIT-PA). A pharmacokinetic screening in clinical urine samples has been also carried out.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16814307 DOI: 10.1016/j.chroma.2006.06.019
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759