Literature DB >> 16810032

Clinical use of indices determined non-invasively from the radial and carotid pressure waveforms.

Audrey Adji1, Kozo Hirata, Michael F O'Rourke.   

Abstract

OBJECTIVE: To evaluate the clinical use of radial and carotid artery applanation tonometry as an independent supplement to cuff sphygmomanometry.
METHODS: In 44 patients, radial and carotid tonometric pressure recordings were taken at short intervals apart by two persons who had prolonged experience with both. Comparisons were made between directly recorded radial and carotid waveforms and between aortic waves synthesized from both, using SphygmoCor. Focus was on waveform features: time intervals between wavefoot and incisura, denoting ejection duration, between wavefoot and first systolic peak or shoulder T1, and augmentation index - the rise in pressure from this point to systolic peak divided by pulse pressure.
RESULTS: No patient had discomfort with radial tonometry, whereas many found carotid tonometry uncomfortable. Beat-to-beat variability was lower for the radial than carotid site. The device's operator "quality index" was achieved for 78% of radial waveforms but just 20% of carotid waveforms (P<0.05). Interobserver variability was lower for all indices derived from radial, cf. carotid, waveforms. For the two observers combined, there was no difference between aortic indices determined from carotid and radial sites except for T1 (radial-derived 117+ or -17 ms, cf. carotid-derived 103+ or -17 ms, P<0.05), but this did not influence the value of augmentation index (radial-derived 26+ or -13%, cf. carotid-derived 28+ or -14%, P=NS).
CONCLUSION: The present study conforms with most published results, and indicates superiority of radial to carotid tonometry in clinical practice.

Entities:  

Mesh:

Year:  2006        PMID: 16810032     DOI: 10.1097/01.mbp.0000218001.50333.b7

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  8 in total

Review 1.  Meta-analysis of the comparative effects of different classes of antihypertensive agents on brachial and central systolic blood pressure, and augmentation index.

Authors:  Charlotte H Manisty; Alun D Hughes
Journal:  Br J Clin Pharmacol       Date:  2013-01       Impact factor: 4.335

Review 2.  Noninvasive measurement of central vascular pressures with arterial tonometry: clinical revival of the pulse pressure waveform?

Authors:  Matthew R Nelson; Jan Stepanek; Michael Cevette; Michael Covalciuc; R Todd Hurst; A Jamil Tajik
Journal:  Mayo Clin Proc       Date:  2010-05       Impact factor: 7.616

3.  Continuous blood pressure monitoring via non-invasive radial artery applanation tonometry and invasive arterial catheter demonstrates good agreement in patients undergoing colon carcinoma surgery.

Authors:  Jing Sun; Hanjian Chen; Jun Zheng; Bin Mao; Shengmei Zhu; Jingyi Feng
Journal:  J Clin Monit Comput       Date:  2016-12-20       Impact factor: 2.502

4.  Relationship between 24-h urine sodium/potassium ratio and central aortic systolic blood pressure in hypertensive patients.

Authors:  Moo-Yong Rhee; Sung-Joon Shin; Namyi Gu; Deuk-Young Nah; Byong-Kyu Kim; Kyung-Soon Hong; Eun-Joo Cho; Ki-Chul Sung; Sim-Yeol Lee; Kwang-Il Kim
Journal:  Hypertens Res       Date:  2016-11-24       Impact factor: 3.872

5.  Epidermis-Inspired Wearable Piezoresistive Pressure Sensors Using Reduced Graphene Oxide Self-Wrapped Copper Nanowire Networks.

Authors:  Yangzhi Zhu; Martin C Hartel; Ning Yu; Pamela Rosario Garrido; Sanggon Kim; Junmin Lee; Praveen Bandaru; Shenghan Guan; Haisong Lin; Sam Emaminejad; Natan Roberto de Barros; Samad Ahadian; Han-Jun Kim; Wujin Sun; Vadim Jucaud; Mehmet R Dokmeci; Paul S Weiss; Ruoxue Yan; Ali Khademhosseini
Journal:  Small Methods       Date:  2021-12-15

6.  Quantitative Comparison of the Performance of Piezoresistive, Piezoelectric, Acceleration, and Optical Pulse Wave Sensors.

Authors:  Hongju Wang; Lu Wang; Nannan Sun; Yang Yao; Liling Hao; Lisheng Xu; Stephen E Greenwald
Journal:  Front Physiol       Date:  2020-01-14       Impact factor: 4.566

7.  Ageing, hypertension and aortic valve stenosis - Understanding the series circuit using cardiac magnetic resonance and applanation tonometry.

Authors:  S L Hungerford; A I Adji; N K Bart; L Lin; N Song; A Jabbour; M F O'Rourke; C S Hayward; D W M Muller
Journal:  Int J Cardiol Hypertens       Date:  2021-05-28

8.  Machine learning for detection of stenoses and aneurysms: application in a physiologically realistic virtual patient database.

Authors:  G Jones; J Parr; P Nithiarasu; S Pant
Journal:  Biomech Model Mechanobiol       Date:  2021-07-31
  8 in total