| Literature DB >> 16803669 |
Thomas Grobosch1, Boris Angelow, Lena Schönberg, Dagmar Lampe.
Abstract
A 55-year-old man came to the hospital with a bleeding wound on his tongue. The coating of his tongue was green, and his sputum was red. Because an increased international normalized ratio-value was measured, a blood sample was sent to our laboratory with the suspicion of coumarin intoxication. Liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS) analysis confirmed the poisoning was by bromadiolone, with its maximum serum concentration at 440 microg/L. The analysis of further samples resulted in a calculated elimination half-life of 140 h. The analytical method described was developed for the determination and quantitation of bromadialone using LC-MS. This method is suitable for the simultaneous identification and quantitation of 10 indirect anticoagulants in human serum, which include five superwarfarins (brodifacoum, bromadiolone, difenacoum, difethialone, and flocoumafen) as rodenticides licenced in Germany and five other vitamin K antagonists (acenocoumarol, coumatetralyl, coumachlor, phenprocoumon, and warfarin). The method is based on an acidic (pH 4.2) liquid-liquid extraction followed by LC-ESI-MS analysis. Analytical separation was carried out using an Atlantis C18 column (2.1 x 20 mm, 3 microm). The mobile phase consisted of methanol/0.1% formic acid; the flow rate was 0.6 mL/min, and the time needed for analysis was 5 min. The lower limit of quantitation was 5 microg/L (signal-to-noise > 10).Entities:
Mesh:
Substances:
Year: 2006 PMID: 16803669 DOI: 10.1093/jat/30.4.281
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367