Gastrointestinal endoscopes cleaned without detergent substance following an automated endoscope washer/disinfector dysfunction.

Marie Méan1, Marie Reine Mallaret, Philippe Bichard, Jacqueline Shum, Jean-Pierre Zarski.   

Abstract

OBJECTIVE: To report cases of gastrointestinal endoscopies performed with endoscopes that were reprocessed without detergent substance during a period of dysfunction of the automated endoscope reprocessor (AER).
METHOD: A dysfunction of the AER for the cycles requiring detergent substance was reported at the Grenoble University Hospital on March 2005. During this period, 72 patients had potentially been exposed to a contaminated endoscope. A recall procedure was organized and serologic tests (HIV, HCV, HBV) were performed 3 and 6 months after the AER incident.
RESULTS: Within the 72 patients convened, 56 (77.8%) were seen in consultation and accepted the serologic screening. Finally, serologic screening was done for 59 patients (81.9%) and no seroconversion for HIV, HCV, or HBV was observed. The final attrition rate was 13 patients (18.1%).
CONCLUSION: No viral infection was transmitted during the AER dysfunction. After this AER incident, the monitoring of the endoscopic procedures and the traceability of the cleaning process were both improved to prevent further incidents.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16801888     DOI: 10.1016/s0399-8320(06)73258-4

Source DB:  PubMed          Journal:  Gastroenterol Clin Biol        ISSN: 0399-8320


  3 in total

Review 1.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

2.  EVOTECH endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy.

Authors:  Michelle J Alfa; Pat DeGagne; Nancy Olson; Iram Fatima
Journal:  BMC Infect Dis       Date:  2010-07-09       Impact factor: 3.090

Review 3.  Infectious diseases linked to cross-contamination of flexible endoscopes.

Authors:  Nikki Kenters; Elisabeth G W Huijskens; Corianne Meier; Andreas Voss
Journal:  Endosc Int Open       Date:  2015-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.