Magnetic resonance-guided focused ultrasound surgery of breast cancer: reliability and effectiveness.

Hidemi Furusawa1, Kiyoshi Namba, Sharon Thomsen, Futoshi Akiyama, Achiude Bendet, Chiaki Tanaka, Yukiko Yasuda, Hiroshi Nakahara.   

Abstract

BACKGROUND: Magnetic resonance-guided focused ultrasound surgery (MRgFUS) is a noninvasive technique that has been shown to coagulate benign and malignant tumors. The purpose of this study was to evaluate MRgFUS safety and effectiveness for the ablation of breast carcinomas. STUDY
DESIGN: Thirty women with biopsy-proved breast cancer underwent MRgFUS treatment. Gadolinium-enhanced MR images were used for treatment planning and posttreatment radiologic assessment of treated tissue, and temperature-sensitive MR images provided real-time treatment monitoring. After MRgFUS, all 30 women underwent wide excision or mastectomy. The extent of thermal ablation was assessed with tumor histology.
RESULTS: Treatment was well tolerated, with a minimum of adverse effects, especially when performed under local anesthesia. On pathologic examination, mean (+/-SD) necrosis of the targeted breast tumors was 96.9 +/- 4% (median 100%, range 78% to 100%) of tumor volume. Fifteen (53.5%) of 28 evaluable patients had 100% necrosis of the ablated tumor; only 3 patients (10.7%) had less than 95% necrosis. In 28 (93.3%) patients, 100% of the malignancy was within the treatment field, and 98% and 95% of tumor lay within the treatment field in 2 remaining patients. Retrospective analysis in two patients with residual tumor showed treatment was not delivered to the full recommended area, reaffirming the need for precise localization and the value of contrast-enhanced images for treatment planning.
CONCLUSIONS: MRgFUS has great potential to become a viable noninvasive replacement for lumpectomy. Additional studies focusing on posttreatment image-based evaluation are needed.

Entities:  

Mesh:

Year:  2006        PMID: 16798487     DOI: 10.1016/j.jamcollsurg.2006.04.002

Source DB:  PubMed          Journal:  J Am Coll Surg        ISSN: 1072-7515            Impact factor:   6.113


  62 in total

1.  Design and characterization of a laterally mounted phased-array transducer breast-specific MRgHIFU device with integrated 11-channel receiver array.

Authors:  A Payne; R Merrill; E Minalga; U Vyas; J de Bever; N Todd; R Hadley; E Dumont; L Neumayer; D Christensen; R Roemer; D Parker
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

Review 2.  High-Intensity Focused Ultrasound: Current Status for Image-Guided Therapy.

Authors:  Alexander Copelan; Jason Hartman; Monzer Chehab; Aradhana M Venkatesan
Journal:  Semin Intervent Radiol       Date:  2015-12       Impact factor: 1.513

Review 3.  MRI-guided laparoscopic and robotic surgery for malignancies.

Authors:  Makoto Hashizume
Journal:  Int J Clin Oncol       Date:  2007-04-27       Impact factor: 3.402

Review 4.  MR thermometry for monitoring tumor ablation.

Authors:  Baudouin Denis de Senneville; Charles Mougenot; Bruno Quesson; Iulius Dragonu; Nicolas Grenier; Chrit T W Moonen
Journal:  Eur Radiol       Date:  2007-05-22       Impact factor: 5.315

5.  MRI-compatible positioning device for guiding a focused ultrasound system for transrectal treatment of prostate cancer.

Authors:  Christos Yiallouras; Nicos Mylonas; Christakis Damianou
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-13       Impact factor: 2.924

6.  [Image-guided therapy with closed MRI: an update].

Authors:  R Hoffmann; H Rempp; P L Pereira; S Clasen
Journal:  Radiologe       Date:  2013-11       Impact factor: 0.635

Review 7.  Applications of Focused Ultrasound in Cerebrovascular Diseases and Brain Tumors.

Authors:  Francesco Prada; M Yashar S Kalani; Kaan Yagmurlu; Pedro Norat; Massimiliano Del Bene; Francesco DiMeco; Neal F Kassell
Journal:  Neurotherapeutics       Date:  2019-01       Impact factor: 7.620

Review 8.  MRI-guided focused ultrasound surgery.

Authors:  Ferenc A Jolesz
Journal:  Annu Rev Med       Date:  2009       Impact factor: 13.739

9.  Multi-parametric monitoring and assessment of high-intensity focused ultrasound (HIFU) boiling by harmonic motion imaging for focused ultrasound (HMIFU): an ex vivo feasibility study.

Authors:  Gary Y Hou; Fabrice Marquet; Shutao Wang; Elisa E Konofagou
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

Review 10.  Update on Clinical Magnetic Resonance-Guided Focused Ultrasound Applications.

Authors:  Thiele Kobus; Nathan McDannold
Journal:  Magn Reson Imaging Clin N Am       Date:  2015-07-07       Impact factor: 2.266

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.