Experimental study of one-shot vascular anastomostic device for proximal vein graft anastomoses.

Luqi Liu1, Junsheng Liu, Meijia Zhu, Shouxiang Hu.   

Abstract

PURPOSE: A new type of sutureless aortic-vein-graft vascular anastomostic device, One-Shot Vascular Anastomostic Device (Horologe Factory of Jinan City, Shandong Province, China) has been recently designed to create a one-shot anastomosis between the aorta and vein grafts for coronary artery bypass grafting surgery. DESCRIPTION: Twelve pigs were scheduled for the test of the feasibility of the One-Shot Vascular Anastomostic Device for artery to vein graft anastomosis. In each animal one proximal anastomosis was performed by means of the One-Shot Vascular Anastomostic Device and the distal end was sutured in a conventional manner to serve as the animal own control. The anastomosis incorporating the abdominal main artery to the segment of a free external carotid vein to the external iliac artery is for the simulation of the aorta-vein anastomosis. EVALUATION: The mean duration of the completion of the proximal anastomosis was 1.2 +/- 1.2 minutes (range, 0.3-5.0 minutes). There was a significant difference between the control groups (p < 0.01). All vein grafts were still functioning at the end of the procedure. Pathologic studies and angiography demonstrated that the results were satisfactory.
CONCLUSIONS: This device enables rapid and safe completion of vascular anastomostic procedure. The vein graft was functioning well and incorporated into the vessel intima smoothly.

Entities:  

Mesh:

Year:  2006        PMID: 16798234     DOI: 10.1016/j.athoracsur.2006.01.099

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  5 in total

1.  Rapid prototyped sutureless anastomosis device from self-curing silk bio-ink.

Authors:  Rod R Jose; Waseem K Raja; Ahmed M S Ibrahim; Pieter G L Koolen; Kuylhee Kim; Abdurrahman Abdurrob; Jonathan A Kluge; Samuel J Lin; Gillian Beamer; David L Kaplan
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2014-11-11       Impact factor: 3.368

2.  Vascular anastomosis using controlled phase transitions in poloxamer gels.

Authors:  Edward I Chang; Michael G Galvez; Jason P Glotzbach; Cynthia D Hamou; Samyra El-ftesi; C Travis Rappleye; Kristin-Maria Sommer; Jayakumar Rajadas; Oscar J Abilez; Gerald G Fuller; Michael T Longaker; Geoffrey C Gurtner
Journal:  Nat Med       Date:  2011-08-28       Impact factor: 53.440

3.  Evaluation of a novel thermosensitive heparin-poloxamer hydrogel for improving vascular anastomosis quality and safety in a rabbit model.

Authors:  Ying-Zheng Zhao; Hai-Feng Lv; Cui-Tao Lu; Li-Juan Chen; Min Lin; Ming Zhang; Xi Jiang; Xiao-Tong Shen; Rong-Rong Jin; Jun Cai; Xin-Qiao Tian; Ho Lun Wong
Journal:  PLoS One       Date:  2013-08-27       Impact factor: 3.240

4.  A strongly adhesive hemostatic hydrogel for the repair of arterial and heart bleeds.

Authors:  Yi Hong; Feifei Zhou; Yujie Hua; Xianzhu Zhang; Chengyao Ni; Dihao Pan; Yiqing Zhang; Deming Jiang; Long Yang; Qiuning Lin; Yiwei Zou; Dongsheng Yu; David E Arnot; Xiaohui Zou; Linyong Zhu; Shufang Zhang; Hongwei Ouyang
Journal:  Nat Commun       Date:  2019-05-14       Impact factor: 14.919

5.  A Clampless Aortoprosthetic End to Side Anastomotic Device with Large Diameter Aortic Puncher.

Authors:  Tadahiro Sasajima; Yukihiro Saito; Hayato Ise; Daiki Uchida; Hiroyuki Kamiya
Journal:  EJVES Vasc Forum       Date:  2020-02-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.