| Literature DB >> 16784831 |
Yunyun Jiang1, Mei Lin, Guorong Fan, Yi Chen, Zhen Li, Weiquan Zhao, Yutian Wu, Jinhong Hu.
Abstract
A simple, sensitive and rapid method for analysis of granisetron in human plasma, utilizing liquid chromatography tandem mass spectrometry (LC-MS/MS), has been developed and validated to satisfy FDA guidelines for bioanalytical methods. The analyte and internal standard (IS) were isolated from 100microl plasma samples by liquid-liquid extraction (LLE). A Varian 1200l tandem mass spectrometer equipped with an electrospray ionization source was operated in selected reaction monitoring (SRM) mode with the precursor-to-product ion transitions m/z 313.4/138 for granisetron and m/z 270/201 for the IS used for quantitation. The assay exhibited a linear dynamic range of 0.02-20ng/ml for granisetron in human plasma. The lower limit of quantification (LLOQ) was 0.02ng/ml with a relative standard deviation of less than 15%. The mean extraction recovery from spiked plasma samples was 97.9%. The intra-day accuracy of the assay was within 10% of nominal and intra-day precision was better than 15% C.V. A run time of 2.0min for each sample made it possible for high-throughput bioanalysis. The method was employed in a bioequivalence study of two formulations of granisetron hydrochloride 1mg rapidly disintegrating tablets/1mg capsules.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16784831 DOI: 10.1016/j.jpba.2006.05.001
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935