Literature DB >> 16784632

[An analysis of failure factors in treatment of femoral shaft fracture with interlocking intramedullary nailing].

Guang-hui Li1, Feng Li, Ren-yun Xia, Ti-pei Wang, An-min Chen.   

Abstract

OBJECTIVE: To analyze the failure factors treating femoral shaft fracture with interlocking intramedullary nailing, and to offer the measures for prevention.
METHODS: From April 1999 to September 2002, among the 213 cases of femoral shaft fracture treated by interlocking intramedullary nailing, 18 were failure. The data of these cases were analyzed.
RESULTS: The failure cases were divided into 6 types: 3 cases with nail break-down, 7 with distal screw backing out, 3 with proximal screw backing out, 3 with fibrous ankylosis of knee, 2 with late infection. Failure relative factors were regarded to operation technique, incorrect rehabilitation choice of operative indication, quality of interlocking intramedullary nailing.
CONCLUSIONS: An appropriate indication should be chosen as using interlocking intramedullary nailing for femoral shaft fracture. The skills should be mastered well. The suitable materials should be chosen carefully. And the rehabilitation exercises should be directed properly. All these are important measures for prevention.

Entities:  

Mesh:

Year:  2006        PMID: 16784632

Source DB:  PubMed          Journal:  Zhonghua Wai Ke Za Zhi        ISSN: 0529-5815


  1 in total

1.  Effective Treatment for Breakage of a Solid Long Cementless Stem with a Distal Locking Screw: A Case Report.

Authors:  Shun Igarashi; Hiroaki Kijima; Shin Yamada; Yosuke Iwamoto; Masashi Fujii; Naohisa Miyakoshi
Journal:  J Orthop Case Rep       Date:  2021-11
  1 in total