Comparability of point-of-care whole-blood electrolyte and substrate testing using a Stat Profile Critical Care Xpress analyzer and standard laboratory methods.

Zlata Flegar-Mestrić1, Sonja Perkov.   

Abstract

BACKGROUND: Rapid technological progress in point-of-care testing allows the measurement of multiple analytes in whole-blood samples. The present study evaluated biosensor-based methods for the measurement of electrolytes and substrates in whole blood using a Stat Profile Critical Care Xpress (Nova Biomedical, Waltham, MA, USA) multiprofile analyzer and their comparability with standard laboratory methods. Because of the increased utilization of arterial blood samples in hospitalized patients and limited information on differences between arterial and venous blood for most routine laboratory tests, analytical differences caused by different sample types were evaluated.
METHODS: Whole-blood arterial samples and venous serum samples were obtained from 70 random patients with a variety of diagnoses admitted to the intensive care unit. The Stat Profile Critical Care Xpress analyzer was used to obtain whole-blood electrolyte and substrate profiles. For comparison studies, plasma or serum samples were analyzed according to standard laboratory methods using an Olympus AU 600 analyzer (Olympus Mishima, Shizuoka, Japan).
RESULTS: Imprecision, expressed as the coefficient of variation (CV%), was less than 5.7% for all analytes at both high and low concentrations, except for creatinine, with a CV of 13.8% for low and 9.5% for high concentrations. The inaccuracy of electrolyte and substrate measurements in whole blood using a Stat Profile Critical Care Xpress analyzer met the analytical quality specification required for near patient testing, with observed bias within the range -4.5% to 5.3%. Statistically significant correlation (p<0.05) was obtained between standard laboratory methods performed on arterial plasma or venous serum samples on an Olympus AU 600 analyzer and direct whole-blood measurements on the Stat Profile Critical Care Xpress point-of-care analyzer for all parameters tested, although slope and intercept values showed analytical differences for electrolyte measurement.
CONCLUSIONS: The Stat Profile Critical Care Xpress multiprofile point-of-care analyzer provides rapid and accurate direct whole-blood measurement with acceptable performance compared to standard laboratory methods. The results obtained for electrolytes and substrates in whole blood were comparable to those for standard laboratory methods using arterial plasma or venous serum samples.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16776641     DOI: 10.1515/CCLM.2006.148

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  9 in total

1.  Agreement of arterial sodium and arterial potassium levels with venous sodium and venous potassium in patients admitted to intensive care unit.

Authors:  Sunil Kumar Nanda; Lopamudra Ray; Asha Dinakaran
Journal:  J Clin Diagn Res       Date:  2015-02-01

2.  Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes.

Authors:  Anunaya Jain; Imron Subhan; Mahesh Joshi
Journal:  Int J Emerg Med       Date:  2009-02-24

3.  Point of care blood gases with electrolytes and lactates in adult emergencies.

Authors:  Dheeraj Kapoor; Meghana Srivastava; Pritam Singh
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

4.  Bedside ABG, electrolytes, lactate and procalcitonin in emergency pediatrics.

Authors:  Prerna Batra; Ajeet Kumar Dwivedi; Neha Thakur
Journal:  Int J Crit Illn Inj Sci       Date:  2014-07

5.  Are sodium and potassium results on arterial blood gas analyzer equivalent to those on electrolyte analyzer?

Authors:  Shalini Gupta; Ashwani K Gupta; Kamaljit Singh; Minni Verma
Journal:  Indian J Crit Care Med       Date:  2016-04

6.  Role of Correction Factor in Minimizing Errors While Calculating Electrolyte Values between Blood-gas Analyzer and Laboratory Autoanalyzer: A Comparative Study.

Authors:  Abhinav Banerjee; Gesu Mehrotra
Journal:  Indian J Crit Care Med       Date:  2018-01

7.  Interchangeability of Sodium and Potassium Result Values of Arterial Blood Gas with Laboratory Analyzer: Narrative Review.

Authors:  Yasemin Ustundağ; Kağan Huysal; Şeyda E Ozgunay; Ali R Turkoğlu
Journal:  Indian J Crit Care Med       Date:  2019-01

8.  Central Venous Blood Gas Analysis: An Alternative to Arterial Blood Gas Analysis for pH, PCO2, Bicarbonate, Sodium, Potassium and Chloride in the Intensive Care Unit Patients.

Authors:  Mubina Begum Bijapur; Nazeer Ahmed Kudligi; Shaik Asma
Journal:  Indian J Crit Care Med       Date:  2019-06

9.  Development and evaluation of three mortality prediction indices for cold-stunned Kemp's ridley sea turtles (Lepidochelys kempii).

Authors:  N I Stacy; C J Innis; J A Hernandez
Journal:  Conserv Physiol       Date:  2013-04-19       Impact factor: 3.079

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.