Novel pachometry calibration.

Amir M Moezzi1, Sokpheaktra Sin, Trefford L Simpson.   

Abstract

PURPOSE: The purpose of this study was to develop a simple method for cross-calibrating instruments that measure corneal thickness.
METHODS: Fourteen rigid lenses of different thicknesses were manufactured using a material with refractive index of 1.376. Center thickness of the lenses (CT) was measured using a computerized optical pachometer (OP), two optical coherence tomographers (OCTs), and a confocal microscope (CM). Accuracy of measurements was compared between the four instruments.
RESULTS: Before calibrating the machines, there was a significant effect of the measurement device (p < 0.05). The differences between instruments were eliminated (p > 0.05) after applying calibration equations for each device. In addition, after each instrument was calibrated with lenses of 1.376 refractive index, there was no significant difference (p > 0.05) between measured values of lens center thickness by OP, each OCT, CM, and the physical center thickness of the lenses.
CONCLUSIONS: Using calibration lenses with the same refractive index as the cornea (1.376) allows rapid and simple calibration of the pachometers so that corneal thickness measurements from different devices can be used interchangeably.

Mesh:

Year:  2006        PMID: 16772888     DOI: 10.1097/01.opx.0000221403.32040.17

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  3 in total

1.  Vertical and horizontal corneal epithelial thickness profiles determined by ultrahigh resolution optical coherence tomography.

Authors:  Chixin Du; Jianhua Wang; Lele Cui; Meixiao Shen; Yimin Yuan
Journal:  Cornea       Date:  2012-09       Impact factor: 2.651

2.  Corneal Sublayers Thickness Estimation Obtained by High-Resolution FD-OCT.

Authors:  Diego Alberto; Roberto Garello
Journal:  Int J Biomed Imaging       Date:  2013-06-02

3.  Corneal Swelling with Cosmetic etafilcon A Lenses versus No Lens Wear.

Authors:  Amir M Moezzi; Jalaiah Varikooty; Marc Schulze; William Ngo; Kathrine Osborn Lorenz; Danielle Boree; Lyndon W Jones
Journal:  Optom Vis Sci       Date:  2016-06       Impact factor: 1.973

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.