Comparing real-world advantages for the clinical neuroradiologist between a high field (3 T), a phased array (1.5 T) vs. a single-channel 1.5-T MR system.

Darren B Orbach1, Chris Wu, Meng Law, James S Babb, Ray Lee, Abraham Padua, Edmond A Knopp.   

Abstract

PURPOSE: To evaluate signal-to-noise ratio (SNR) and neuroradiologists' subjective assessments of image quality in 3-Tesla (3-T) or phased-array MR systems that are now available for clinical neuroimaging.
MATERIALS AND METHODS: Brain MR images of six normal volunteers were obtained on each of three scanners: a 1.5-T single-channel system, a 12-channel, phased-array system, and a 3-T single-channel system. Additionally, clinically optimized images acquired from 28 patients who underwent imaging in more than one of these systems were analyzed. SNRs were measured and image quality and artifact conspicuity were graded by two blinded readers.
RESULTS: The phased-array system produced higher SNR than either the 1.5-T or the 3-T single-channel systems, and in no instance was it outperformed. Both blinded readers judged the phased-array images to be of higher quality than those produced by the single-channel systems, with significantly less artifact. The 3-T magnet produced images with high SNR, but with increased artifact conspicuity. The phased-array system markedly decreased acquisition times without introduction of artifacts.
CONCLUSION: Both quantitatively and qualitatively, the phased-array system provided image quality superior to that of the 1.5-T and 3-T single-channel systems. (c) 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2006        PMID: 16767701     DOI: 10.1002/jmri.20612

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  1 in total

1.  Hyperintense acute reperfusion marker on FLAIR is not associated with early haemorrhagic transformation in the elderly.

Authors:  Michal Rozanski; Martin Ebinger; Wolf U Schmidt; Benjamin Hotter; Sandra Pittl; Peter U Heuschmann; Jan G Jungehuelsing; Jochen B Fiebach
Journal:  Eur Radiol       Date:  2010-07-21       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.