| Literature DB >> 16765108 |
Juan Manuel Serrano1, Manuel Silva.
Abstract
A simple and sensitive method was developed for the quantification of amikacin in human plasma and urine samples. The method involves centrifugation of body fluid plasma after dilution with an ethanol/sodium carbonate mixture, and then an aliquot of the supernatant is directly injected into the chromatograph. After separation on a reversed-phase C18 column (runtime 20 min), aminoglycoside is detected on the basis of its complex formation reaction with Cu(II), the catalyst of the luminol/hydrogen peroxide chemiluminescence system. Using a volume of 500 microl biological sample, linearity is established over the concentration range 0.15-2.0 microg/ml and the limit of detection (LOD) is ca. 50 microg/l in plasma or urine. The intra-day and inter-day precision (measured by relative standard deviation, R.S.D.%) are always less than 9%, and relative recoveries are found to be over 92%.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16765108 DOI: 10.1016/j.jchromb.2006.05.016
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205