Literature DB >> 16762998

Performance of a new-generation chemiluminescent assay for hepatitis B surface antigen.

Dan Chen1, Lawrence A Kaplan.   

Abstract

BACKGROUND: The usual criteria for analysis of hepatitis B surface antigen (HBsAg) are detection of HBsAg and result confirmation by antibody neutralization. We observed that with the Immulite 2000 HBsAg assay [Diagnostics Product Corporation (DPC)] a relatively high percentage of weakly reactive (WR) samples did not pass the neutralization step.
METHODS: For each of 3 lots of Immulite 2000 HBsAg reagent (DPC), we collected and analyzed HBsAg data from approximately 3000 to 4000 patient blood samples and compared these data with HBsAg data from 3393 samples tested with the Abbott Auszyme assay. For 127 samples with initially WR detection signals (relative signal/cutoff index of 1.00-2.5) on the Immulite 2000 HBsAg assay, we then measured hepatitis B (HB) viral load and/or other HB serologic markers.
RESULTS: The Immulite 2000 HBsAg assay produced more initially reactive results than the Abbott Auszyme method. Many of these reactive samples, however, were WR and did not meet the confirmation criteria in the neutralization test. Moreover, DNA PCR testing indicated that 22 of the 38 WR samples (58%) that did meet the confirmation criteria had no detectable HB viral DNA.
CONCLUSIONS: Immulite 2000 HBsAg assay results include a unique group of WR samples that are associated with both false-positive and false-negative results, regardless of neutralization status, and require careful interpretation. WR HBsAg samples should be reported as confirmed HBsAg reactive only if the samples not only meet the neutralization criteria but also are positive for other HB serologic markers such as anti-HB core total and anti-HB core IgM.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16762998     DOI: 10.1373/clinchem.2005.064063

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

Review 1.  The underlying mechanisms for the "isolated positivity for the hepatitis B surface antigen (HBsAg)" serological profile.

Authors:  Robério Amorim de Almeida Pondé
Journal:  Med Microbiol Immunol       Date:  2010-05-11       Impact factor: 3.402

2.  Impact of grey zone sample testing by enzyme-linked immunosorbent assay in enhancing blood safety: Experience at a tertiary care hospital in North India.

Authors:  Archana Solanki; Abhay Singh; Rajendra Chaudhary
Journal:  Asian J Transfus Sci       Date:  2016 Jan-Jun

3.  Diagnostic validity of the chemiluminescent method compared to polymerase chain reaction for hepatitis B virus detection in the routine clinical diagnostic laboratory.

Authors:  Mohammad-Hassan Khadem-Ansari; Mir-Davood Omrani; Yousef Rasmi; Arsalan Ghavam
Journal:  Adv Biomed Res       Date:  2014-05-28