Use of intra-cardiac ultrasound in the diagnosis of prosthetic valve malfunction.

Faizel Osman1, Rick Steeds.   

Abstract

A 46-year old lady was under regular follow up for aortic valve replacement done in 2002 for aortic stenosis. The valve was a Carbomedics 19 mm bi-leaflet aortic valve prosthesis for which she had adequate anticoagulation since implantation. She had a past history of end stage renal failure, type 2 diabetes, hypertension, cerebrovascular disease and systemic lupus erythematosus. She was asymptomatic and had a routine transthoracic echocardiogram performed which revealed that her aortic valve prosthesis was well seated with an elevated velocity across the valve; the aortic valve leaflets were poorly visualised but appeared to be mobile. A transoesophageal echocardiogram (TOE) confirmed the markedly increased forward velocity across the aortic valve (maximum velocity 4.6 m/s) with small aortic root raising the possibility of pressure recovery phenomenon. Once again the leaflets were not clearly seen but appeared mobile despite the use of deep trans-gastric view. Fluoroscopy was performed and revealed that one of the leaflets was not moving. The patient had an intra-cardiac ultrasound scan (ACUSON AcuNav; Siemens) with the probe of the scanner within the right atrium. A long-axis view demonstrated the prosthetic aortic valve leaflets clearly (Fig. 1). A short-axis view of the prosthetic valve revealed an echogenic area at the six o'clock position; this may be due to pannus formation (Fig. 2); colour flow across the valve during systole revealed absence of colour flow though one of the leaflets due to the leaflet being stuck (Fig. 3). The sewing cuff of the Carbomedics valve is coated with biolite carbon, which is an anti-thrombotic agent that prevents adhesion of thrombus or pannus on the sewing cuff. There are few reports of Carbomedics valve dysfunction by pannus formation in the mitral position but none in the aortic position. Fluoroscopy can be used to visualize mobility of valve leaflets but is unable to identify thrombus/pannus formation that may be causing the valve to stick. Intra-cardiac echocardiography (ICE) can provide additional information regarding potential causes of valve dysfunction such as pannus/thrombus formation. However, ICE does suffer from the same limitations as TOE in obtaining sufficiently adequate images of the aortic valve to allow appropriate assessment of leaflet motion and insight into flow channels; it is also more invasive and expensive compared with other imaging modalities. In our patient ICE provided better images than TOE, possibly due to the closer proximity of the probe to the aortic root compared with TOE. ICE is a novel tool in the assessment of prosthetic valve function, which can complement data obtained from transthoracic/transoesophageal echocardiography and fluoroscopy.

Entities:  

Mesh:

Year:  2006        PMID: 16762598     DOI: 10.1016/j.euje.2006.04.002

Source DB:  PubMed          Journal:  Eur J Echocardiogr        ISSN: 1532-2114


  5 in total

1.  Intracardiac echocardiography to diagnose pannus formation after aortic valve replacement.

Authors:  Yoshiya Yamamoto; Takahiro Ohara; Akira Funada; Hiroyuki Takahama; Makoto Amaki; Takuya Hasegawa; Yasuo Sugano; Hideaki Kanzaki; Toshihisa Anzai
Journal:  J Echocardiogr       Date:  2016-01-05

Review 2.  Intracardiac Echocardiography for Structural Heart and Electrophysiological Interventions.

Authors:  Craig Basman; Yuvrajsinh J Parmar; Itzhak Kronzon
Journal:  Curr Cardiol Rep       Date:  2017-09-06       Impact factor: 2.931

3.  Intraprocedural guidance: which imaging technique ranks highest and which one is complementary for closing paravalvular leaks?

Authors:  Thomas Bartel; Silvana Müller
Journal:  Cardiovasc Diagn Ther       Date:  2014-08

Review 4.  Why is intracardiac echocardiography helpful? Benefits, costs, and how to learn.

Authors:  Thomas Bartel; Silvana Müller; Angelo Biviano; Rebecca T Hahn
Journal:  Eur Heart J       Date:  2013-10-21       Impact factor: 29.983

Review 5.  Recent advances in echocardiography for valvular heart disease.

Authors:  Rebecca Hahn
Journal:  F1000Res       Date:  2015-09-28
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.