Intercomparison of dosimetry systems based on CaF2:Mn TL detectors.

B Zorko1, S Miljanić, B Vekić, M Stuhec, S Gobec, M Ranogajec-Komor.   

Abstract

The responses of readings by the TL dosimetry system MR200 TL developed in-house and used at JSI and the TOLEDO TL system used at RBI are compared. Ten measurements at different doses ranging from 0.01 mSv to 5 Sv were carried out. A set of 36 dosemeters with three pellets of CaF2:Mn were irradiated in radiation fields of 137Cs and 60Co. Analysis of the measured results shows that at doses below 0.1 Sv, readers' outputs do not differ >5% from each other. At doses >1 Sv, the results obtained by the MR200 reader must be corrected with a known factor. Finally, the reproducibility of the results from the MR200 was tested.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16735564     DOI: 10.1093/rpd/nci640