Comparison of the energy-response factor of LiF and Al2O3 in radiotherapy beams.

Paul Mobit1, Ephraim Agyingi, George Sandison.   

Abstract

A Monte Carlo study of the energy-response factor of aluminium oxide (Al2O3) and lithium fluoride (LiF) TLDs in kilovoltage and megavoltage photon beams relative to 60Co gamma rays has been performed using EGSnrc Monte Carlo simulations. The sensitive volume of the detector was simulated as a disc of diameter 2.85 mm and thickness 1 mm. The phantom material was water and the irradiation depth was 2.0 cm in kilovoltage photon beams and 5.0 cm for megavoltage photon beams. The results show that the energy-response of the Al2O3 and LiF-TLDs is constant within 3% for photon beam energies in the energy range of 60Co gamma rays to 25 MV X rays. However, both detectors show an enhanced response for kilovoltage photon beams, which in the case of 50 kV X rays is 3.2 times higher than that for 60Co gamma rays. The energy-response factor was 1.46 for LiF irradiated in 50 kV X rays. The Al2O3 detector has an energy-response that is 2.2 times higher than that of LiF in 50 kV X rays decreasing to 1.19 for 250 kV X rays. The results show that the addition of 0.1 or 1% of carbon by weight (as dopant) into the Al2O3 does not change the Monte Carlo determined energy-response factor by more than 1%.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16735558     DOI: 10.1093/rpd/nci676

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  2 in total

1.  Characterisation of energy response of Al(2)O(3):C optically stimulated luminescent dosemeters (OSLDs) using cavity theory.

Authors:  S B Scarboro; S F Kry
Journal:  Radiat Prot Dosimetry       Date:  2012-05-31       Impact factor: 0.972

2.  Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons.

Authors:  Wrya Parwaie; Soheila Refahi; Mahdieh Afkhami Ardekani; Bagher Farhood
Journal:  J Med Signals Sens       Date:  2018 Jul-Sep
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.