Magnetic resonance imaging for assessment of radiofrequency lesions in kidney tissue immediately after ablation: an experimental study.

Axel Häcker1, F Risse, Kristina Peters, Christian Fink, Christel Weiss, Peter E Huber, Peter Alken, Maurice Stephan Michel, Jürgen W Jenne.   

Abstract

BACKGROUND AND
PURPOSE: Radiofrequency ablation (RFA) is an attractive minimally invasive treatment option for small renal masses. The purpose of this study was to investigate the morphologic imaging appearance of RF lesions immediately after the ablation of kidney tissue using standard clinical MR sequences, as well as to investigate the correlation between MR and gross lesion size.
MATERIALS AND METHODS: Ablations were performed 17 times in a standardized model of ex-vivo perfused porcine kidneys using a resistance-controlled RF device (250 W, 470 kHz) and a nonexpandable bipolar applicator inserted into the center of healthy renal parenchyma. The RF current was applied for 9 minutes at 20 W. Imaging was performed after ablation using standard clinical MR sequences: morphologic T(1)/T(2)- weighted images and an isotropic post-contrast T(1) high-resolution measurement (VIBE). Maximum lesion diameters were measured in three directions and were compared with the measurements of the gross lesions. Histologic (hematoxylin + eosin and nicotinamide adenine dinucleotide staining) and statistical analyses were performed.
RESULTS: The gross pathologic examination showed a firm, white-yellow ablation zone sharply demarcated from the untreated tissue. The histologic examination confirmed cellular viability outside but not in the treatment zone. The RF lesions were hyperintense on T(1)-weighted images and hypointense on T(2)-weighted images. The lesion size measured in the VIBE images correlated best with the macroscopic lesion size (N = 16).
CONCLUSIONS: Morphologic MR T(1) and T(2) sequences of RF lesions immediately after ablation produce reliable and consistent imaging characteristics. The post-contrast, high-resolution sequence (VIBE) enables the extent of the lesion to be determined accurately. The potential uses of this imaging strategy in clinical practise warrant further investigation on human renal-cell carcinoma.

Entities:  

Mesh:

Year:  2006        PMID: 16724901     DOI: 10.1089/end.2006.20.312

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  1 in total

Review 1.  Radiofrequency thermal ablation of renal tumors.

Authors:  Massimo De Filippo; Francesca Bozzetti; Rosa Martora; Raffaella Zagaria; Stefania Ferretti; Luca Macarini; Luca Brunese; Antonio Rotondo; Cristina Rossi
Journal:  Radiol Med       Date:  2014-07-15       Impact factor: 3.469

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.