Literature DB >> 16723439

The quality of reporting of diagnostic accuracy studies in glaucoma using the Heidelberg retina tomograph.

Manoharan Shunmugam1, Augusto Azuara-Blanco.   

Abstract

PURPOSE: Scanning laser tomography with the Heidelberg retina tomograph (HRT; Heidelberg Engineering, Heidelberg, Germany) has been proposed as a useful diagnostic test for glaucoma. This study was conducted to evaluate the quality of reporting of published studies using the HRT for diagnosing glaucoma.
METHODS: A validated Medline and hand search of English-language articles reporting on measures of diagnostic accuracy of the HRT for glaucoma was performed. Two reviewers selected and appraised the papers independently. The Standards for Reporting of Diagnostic Accuracy (STARD) checklist was used to evaluate the quality of each publication.
RESULTS: A total of 29 articles were included. Interobserver rating agreement was observed in 83% of items (kappa=0.76). The number of STARD items properly reported ranged from 5 to 18. Less than a third of studies (7/29) explicitly reported more than half of the STARD items. Descriptions of key aspects of the methodology were frequently missing. For example, the design of the study (prospective or retrospective) was reported in 6 of 29 studies, and details of participant sampling (e.g., consecutive or random selection) were described in 5 of 29 publications. The commonest description of diagnostic accuracy was sensitivity and specificity (25/29) followed by area under the ROC curve (13/29), with 9 of 29 publications reporting both.
CONCLUSIONS: The quality of reporting of diagnostic accuracy tests for glaucoma with HRT is suboptimal. The STARD initiative may be a useful tool for appraising the strengths and weaknesses of diagnostic accuracy studies.

Entities:  

Mesh:

Year:  2006        PMID: 16723439     DOI: 10.1167/iovs.05-1250

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  3 in total

Review 1.  Sample size in studies on diagnostic accuracy in ophthalmology: a literature survey.

Authors:  Frank Bochmann; Zoe Johnson; Augusto Azuara-Blanco
Journal:  Br J Ophthalmol       Date:  2007-02-14       Impact factor: 4.638

2.  Diagnostic Performance of the ISNT Rule for Glaucoma Based on the Heidelberg Retinal Tomograph.

Authors:  Errol Wei'en Chan; Jiemin Liao; Reuben Chao Ming Foo; Seng Chee Loon; Tin Aung; Tien Yin Wong; Ching-Yu Cheng
Journal:  Transl Vis Sci Technol       Date:  2013-06-28       Impact factor: 3.283

3.  STARD 2015 was reproducible in a large set of studies on glaucoma.

Authors:  Gianni Virgili; Manuele Michelessi; Alba Miele; Francesco Oddone; Giada Crescioli; Valeria Fameli; Ersilia Lucenteforte
Journal:  PLoS One       Date:  2017-10-12       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.