Literature DB >> 16722480

Evaluating microkeratome efficacy by 3D corneal lamellar flap thickness accuracy and reproducibility using Artemis VHF digital ultrasound arc-scanning.

Dan Z Reinstein1, Hugo F S Sutton, Sabong Srivannaboon, Ronald H Silverman, Timothy J Archer, D Jackson Coleman.   

Abstract

PURPOSE: To present a method that determines the efficacy of flap creation using a microkeratome.
METHODS: Thirty-six eyes of 18 patients were included in the study, operated by two experienced surgeons using the Moria LSK-One microkeratome. Very high-frequency (VHF) digital ultrasound arc-scanning using Artemis technology was used to measure the Reinstein Flap Thickness Profile created by addition of the preoperative epithelial thickness profile to the postoperative stromal component of the flap. Descriptive statistics revealed central flap thickness accuracy and reproducibility. Univariate and multivariate regression analysis was used to determine correlations between central flap thickness and preoperative clinical parameters. Three-dimensional flap thickness maps demonstrated the mean, median, standard deviation, and range of thickness.
RESULTS: Intended flap thickness was 160 microm. Mean central flap thickness was 161 microm in the right eye and 166 microm in the left eye. Central reproducibility (standard deviation of the mean) was 30.3 microm, with a range of 106 to 228 microm. Central flap thickness correlated to preoperative corneal thickness measured by VHF digital ultrasound but not by Orbscan. On qualitative analysis, no consistency among eyes was observed in the three-dimensional flap thickness map. Flaps were on average thinner in the center, but not in all cases.
CONCLUSIONS: A systematic qualitative and quantitative statistical evaluation of microkeratome accuracy and reproducibility can be performed by measuring corneal flap thickness with arc-scanning VHF digital ultrasound.

Entities:  

Mesh:

Year:  2006        PMID: 16722480     DOI: 10.3928/1081-597X-20060501-03

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  4 in total

1.  Artemis very high frequency digital ultrasound-guided femtosecond laser recut after flap complication.

Authors:  Dan Z Reinstein; Zachary Dickeson; Timothy J Archer; Marine Gobbe
Journal:  Digit J Ophthalmol       Date:  2014-09-14

2.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

3.  Rescue of primary incomplete microkeratome flap with secondary femtosecond laser flap in LASIK.

Authors:  E A Razgulyaeva
Journal:  Case Rep Ophthalmol Med       Date:  2014-11-23

4.  Effect of Flat Cornea on Visual Outcome after LASIK.

Authors:  Engy Mohamed Mostafa
Journal:  J Ophthalmol       Date:  2015-11-29       Impact factor: 1.909

  4 in total