Developing pulsatile flow in a deployed coronary stent.

Divakar Rajamohan1, Rupak K Banerjee, Lloyd H Back, Ashraf A Ibrahim, Milind A Jog.   

Abstract

A major consequence of stent implantation is restenosis that occurs due to neointimal formation. This patho-physiologic process of tissue growth may not be completely eliminated. Recent evidence suggests that there are several factors such as geometry and size of vessel, and stent design that alter hemodynamic parameters, including local wall shear stress distributions, all of which influence the restenosis process. The present three-dimensional analysis of developing pulsatile flow in a deployed coronary stent quantifies hemodynamic parameters and illustrates the changes in local wall shear stress distributions and their impact on restenosis. The present model evaluates the effect of entrance flow, where the stent is placed at the entrance region of a branched coronary artery. Stent geometry showed a complex three-dimensional variation of wall shear stress distributions within the stented region. Higher order of magnitude of wall shear stress of 530 dyn/cm2 is observed on the surface of cross-link intersections at the entrance of the stent. A low positive wall shear stress of 10 dyn/cm2 and a negative wall shear stress of -10 dyn/cm2 are seen at the immediate upstream and downstream regions of strut intersections, respectively. Modified oscillatory shear index is calculated which showed persistent recirculation at the downstream region of each strut intersection. The portions of the vessel where there is low and negative wall shear stress may represent locations of thrombus formation and platelet accumulation. The present results indicate that the immediate downstream regions of strut intersections are areas highly susceptible to restenosis, whereas a high shear stress at the strut intersection may cause platelet activation and free emboli formation.

Entities:  

Mesh:

Year:  2006        PMID: 16706584     DOI: 10.1115/1.2194067

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  2 in total

Review 1.  Factors that affect mass transport from drug eluting stents into the artery wall.

Authors:  Barry M O'Connell; Tim M McGloughlin; Michael T Walsh
Journal:  Biomed Eng Online       Date:  2010-03-09       Impact factor: 2.819

2.  Evaluation of the effect of stent strut profile on shear stress distribution using statistical moments.

Authors:  Juan Mejia; Bilal Ruzzeh; Rosaire Mongrain; Richard Leask; Olivier F Bertrand
Journal:  Biomed Eng Online       Date:  2009-04-30       Impact factor: 2.819

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.