Literature DB >> 16705267

Design, in vitro and in vivo assessment of a multi-channel sieve electrode with integrated multiplexer.

Anup Ramachandran1, Martin Schuettler, Natalia Lago, Thomas Doerge, Klaus Peter Koch, Xavier Navarro, Klaus-Peter Hoffmann, Thomas Stieglitz.   

Abstract

This paper reports on the design, in vitro and in vivo investigation of a flexible, lightweight, polyimide based implantable sieve electrode with a hybrid assembly of multiplexers and polymer encapsulation. The integration of multiplexers enables us to connect a large number of electrodes on the sieve using few input connections. The implant assembly of the sieve electrode with the electronic circuitry was verified by impedance measurement. The 27 platinum electrodes of the sieve were coated with platinum black to reduce the electrode impedance. The impedance magnitude of the electrode sites on the sieve (geometric surface area 2,200 microm(2)) was |Z(f=1kHz)| = 5.7 kOmega. The sieve electrodes, encased in silicone, have been implanted in the transected sciatic nerve of rats. Initial experiments showed that axons regenerated through the holes of the sieve and reinnervated distal target organs. Nerve signals were recorded in preliminary tests after 3-7 months post-implantation.

Entities:  

Mesh:

Year:  2006        PMID: 16705267     DOI: 10.1088/1741-2560/3/2/005

Source DB:  PubMed          Journal:  J Neural Eng        ISSN: 1741-2552            Impact factor:   5.379


  6 in total

1.  A hybrid approach for the control of axonal outgrowth: preliminary simulation results.

Authors:  Gianni Ciofani; Pier Nicola Sergi; Jacopo Carpaneto; Silvestro Micera
Journal:  Med Biol Eng Comput       Date:  2010-10-06       Impact factor: 2.602

2.  Microchannel-based regenerative scaffold for chronic peripheral nerve interfacing in amputees.

Authors:  Akhil Srinivasan; Mayank Tahilramani; John T Bentley; Russell K Gore; Daniel C Millard; Vivek J Mukhatyar; Anish Joseph; Adel S Haque; Garrett B Stanley; Arthur W English; Ravi V Bellamkonda
Journal:  Biomaterials       Date:  2014-12-09       Impact factor: 12.479

3.  Functional recordings from awake, behaving rodents through a microchannel based regenerative neural interface.

Authors:  Russell K Gore; Yoonsu Choi; Ravi Bellamkonda; Arthur English
Journal:  J Neural Eng       Date:  2015-01-21       Impact factor: 5.379

4.  Improving impedance of implantable microwire multi-electrode arrays by ultrasonic electroplating of durable platinum black.

Authors:  Sharanya Arcot Desai; John D Rolston; Liang Guo; Steve M Potter
Journal:  Front Neuroeng       Date:  2010-05-06

5.  Flexible active electrode arrays with ASICs that fit inside the rat's spinal canal.

Authors:  Vasiliki Giagka; Andreas Demosthenous; Nick Donaldson
Journal:  Biomed Microdevices       Date:  2015-12       Impact factor: 2.838

6.  Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics.

Authors:  Alejandro Carnicer-Lombarte; Shao-Tuan Chen; George G Malliaras; Damiano G Barone
Journal:  Front Bioeng Biotechnol       Date:  2021-04-15
  6 in total