Literature DB >> 16704592

Simulated clinical evaluation of four fluid warming devices*.

M Turner1, I Hodzovic, W W Mapleson.   

Abstract

The fluid warming capabilities of the Bair Hugger, Hotline, Standard Ranger and Fluido devices were evaluated in the laboratory with gravity flow via a 14G cannula (1 m head of fluid) and with the fluid bag pressurised to 300 mmHg. The resulting flows (70-450 ml.min(-1)) were recorded. At a room temperature of 22 degrees C, no device warmed the fluid to 37 degrees C. However, the Standard Ranger with gravity flow and the Fluido with both gravity and pressurised flow achieved 35 degrees C, whereas the Bair Hugger and Hotline with both gravity and pressurised flow, and the Standard Ranger with pressurised flow, achieved only 24-31 degrees C. However, from the way delivered temperatures changed with flow, we calculated that all four devices would achieve close to 37 degrees C at the flows specified by the manufacturers: 17, 83, 150 and 800 ml.min(-1) for the Bair Hugger, Hotline, Standard Ranger and Fluido, respectively.

Mesh:

Year:  2006        PMID: 16704592     DOI: 10.1111/j.1365-2044.2006.04589.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  7 in total

1.  Effect of a new heated and humidified breathing circuit with a fluid-warming device on intraoperative core temperature: a prospective randomized study.

Authors:  Eugene Kim; Sue-Young Lee; Young-Jin Lim; Jung-Yoon Choi; Young-Tae Jeon; Jung-Won Hwang; Hee-Pyoung Park
Journal:  J Anesth       Date:  2015-03-14       Impact factor: 2.078

2.  Experimental comparison of performances of Mega Acer Kit, Ranger and ThermoSens according to flow rates and distances.

Authors:  Hong Ju Seo; Sang Hun Kim; Tae Hun An; Dong Joon Kim
Journal:  J Clin Monit Comput       Date:  2017-02-07       Impact factor: 2.502

3.  Clinical evaluation of a newly designed fluid warming kit on fluid warming and hypothermia during spinal surgery.

Authors:  Ki Tae Jung; Sang Hun Kim; Keum Young So; Hyeong Jin So; Soo Bin Shim
Journal:  Korean J Anesthesiol       Date:  2015-09-30

4.  Comparing ambient, air-convection, and fluid-convection heating techniques in treating hypothermic burn patients, a clinical RCT.

Authors:  Britt-Marie Kjellman; Mats Fredrikson; Gunilla Glad-Mattsson; Folke Sjöberg; Fredrik Rm Huss
Journal:  Ann Surg Innov Res       Date:  2011-07-07

5.  Can latent heat safely warm blood? - in vitro testing of a portable prototype blood warmer.

Authors:  Mark P McEwen; David Roxby
Journal:  BMC Emerg Med       Date:  2007-08-01

6.  Effect of irrigation fluid temperature on body temperature during arthroscopic elbow surgery in dogs.

Authors:  K R Thompson; P D MacFarlane
Journal:  Open Vet J       Date:  2013-10-19

7.  Mega Acer Kit® is more effective for warming the intravenous fluid than Ranger™ and ThermoSens® at 440 ml/h of infusion rate: an experimental performance study.

Authors:  Dong Joon Kim; Sang Hun Kim; Keum Young So; Tae Hun An
Journal:  Korean J Anesthesiol       Date:  2017-06-14
  7 in total