Validation of the Microlife BPA100 Plus device for self-home blood pressure measurement according to the International Protocol.

George S Stergiou1, Periklis P Giovas, Michalis S Neofytou, Dionysios N Adamopoulos.   

Abstract

OBJECTIVE: Current guidelines recommend that only validated devices for blood pressure measurement should be used. This study presents the validation results of the Microlife BPA100 Plus monitor for self-home blood pressure measurement, according to the European Society of Hypertension International Protocol.
METHODS: Fifteen study participants were included in phase 1 and an additional 18 in phase 2 (total 33). Simultaneous blood pressure measurements were taken by two trained observers (Y tube connected mercury sphygmomanometers) four times, sequentially with three measurements taken using the tested device. Absolute differences between observer and device blood pressure were classified into three zones (within 5, 10 and 15 mmHg). The number of readings with a difference within 5 mmHg was calculated for each individual.
RESULTS: In phase 1, the device produced 32, 42 and 43 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure and 31, 45 and 45 for diastolic blood pressure. In phase 2.1, the device produced 71, 87 and 96 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 71, 98 and 99 measurements for diastolic blood pressure. In phase 2.2, 26 participants had at least two of their differences within 5 mmHg and three participants had no differences within 5 mmHg for systolic blood pressure, whereas for diastolic blood pressure 24 and three participants, respectively. Mean systolic blood pressure differences were -2.0+/-6.0 mmHg and diastolic blood pressure differences were -3.1+/-4.1 mmHg.
CONCLUSIONS: The Microlife BPA100 Plus device for self-home blood pressure measurement comfortably passes the validation requirements of the International Protocol and therefore can be recommended for clinical use in the adult population.

Entities:  

Mesh:

Year:  2006        PMID: 16702824     DOI: 10.1097/01.mbp.0000209071.84965.bf

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  5 in total

1.  Opportunistic screening for hypertension: what does it say about the true epidemiology?

Authors:  Ariadni Menti; Natasa Kalpourtzi; Magda Gavana; Apostolos Vantarakis; Paraskevi V Voulgari; Christos Hadjichristodoulou; Eugenia Gkaliagkousi; Michael Doumas; Rigas G Kalaitzidis; Manolis S Kallistratos; Argiro Karakosta; Vasiliki Katsi; Xenophon Krokidis; Efstathios Manios; Maria Marketou; Angeliki Ntineri; John A Papadakis; Dimitrios Papadopoulos; Pantelis Sarafidis; Grigoris Trypsianis; Michail Chatzopoulos; Grigoris Chlouverakis; Yannis Alamanos; Pantelis Zebekakis; Giota Touloumi; George S Stergiou
Journal:  J Hum Hypertens       Date:  2021-04-09       Impact factor: 3.012

2.  Relevance to home blood pressure monitoring protocol of blood pressure measurements taken before first- morning micturition and in the afternoon.

Authors:  Antonio Eduardo Monteiro de Almeida; Ricardo Stein; Miguel Gus; João Agnaldo Nascimento; Karlyse Claudino Belli; Jorge Rene Garcia Arévalo; Flávio Dani Fuchs; Jorge Pinto Ribeiro
Journal:  Arq Bras Cardiol       Date:  2014-09-23       Impact factor: 2.000

3.  Agreement Between Automated and Human Measurements of Heart Rate in Patients With Atrial Fibrillation.

Authors:  Ting-Tse Lin; Chia-Ling Wang; Min-Tsun Liao; Chao-Lun Lai
Journal:  J Cardiovasc Nurs       Date:  2018 Sep/Oct       Impact factor: 2.083

4.  High prevalence of cardiovascular risk factors in adults living in Greece: the EMENO National Health Examination Survey.

Authors:  Giota Touloumi; Argiro Karakosta; Natasa Kalpourtzi; Magda Gavana; Apostolos Vantarakis; Maria Kantzanou; Christos Hajichristodoulou; Grigoris Chlouverakis; Grigoris Trypsianis; Paraskevi V Voulgari; Yannis Alamanos; Konstantinos Makrilakis; Stavros Liatis; Stylianos Chatzipanagiotou; George Stergiou
Journal:  BMC Public Health       Date:  2020-11-07       Impact factor: 3.295

5.  Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Authors:  Jalil Belghazi; Ramzi N El Feghali; Thérèse Moussalem; Maya Rejdych; Roland G Asmar
Journal:  Vasc Health Risk Manag       Date:  2007
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.