Monitoring the radiation field at aviation altitudes is achieved by the dosemeter system piDOS installed in a passenger aircraft. The basic detector is a 2 in. tissue-equivalent proportional counter (TEPC) mounted in a cabin-baggage-sized aluminium suitcase. The entire system was characterised in neutron and photon reference fields from which two calibration factors were determined according to the splitting of the measured dose deposition spectrum yd(y) into low-LET and high-LET regions. A total of 255 flights in the northern hemisphere was analysed. The dependencies of the ambient dose equivalent rates on altitude, latitude and solar activity have been determined. These new data extend the data base used in Germany for the validation of program codes to calculate the radiation exposure of air crew members.Entities:
Mesh:
Year: 2006 PMID: 16698966 DOI: 10.1093/rpd/ncl053
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972