Statistical analysis of X-ray speckle at the NSLS.

O K Tsui1, S G Mochrie, L E Berman.   

Abstract

A statistical analysis of the static speckle produced by illuminating a disordered aerogel sample by a nominally coherent X-ray beam at wiggler beamline X25 at the National Synchrotron Light Source is reported. The results of the analysis show that the coherence delivered to the X25 hutch is within 35% of what is expected. The rate of coherent photons is approximately two times smaller than expected on the basis of the X25 wiggler source brilliance.

Entities:  

Year:  1998        PMID: 16687798     DOI: 10.1107/S090904959701039X

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616