In vitro validation of right ventricular volume and mass measurement by real-time three-dimensional echocardiography.

Guozhen Chen1, Kun Sun, Guoying Huang.   

Abstract

OBJECTIVE: To evaluate initially the feasibility and accuracy of real-time three-dimensional echocardiography (RT-3DE) for quantifying right ventricular (RV) volume and wall mass in an in vitro experimental study.
METHODS: In ten excised porcine hearts, measurements of RV volume and free wall mass with RT-3DE were outlined and calculated by 2-, 4-, 8- and 16-plane methods with Tom Tec 4D Cardio-View RT 1.0. The results were compared with those of 2D length method and 2D biplane Simpson method. The values of RV silicone latex cast and free wall mass measured by water displacement were served as reference values.
RESULTS: RV shapes of excised porcine hearts with RT-3DE were similar to those of the actual anatomic RVs and RV silicone latex casts. From the findings of analysis of variance and Student-Newman-Keuls test, there was no significant difference between measurements of RV volume with RT-3DE 16-plane (mean 64.05 ml), 8-plane (61.83 ml) and the reference values of RV silicone latex casts (62.94 ml). No significant difference was found between measurements of RV free wall mass with 16-plane (72.81 g), 8-plane (71.05 g) and the reference values of RV free wall masses (76.21 g). However, there was significant difference between measurements of RV volume and free wall mass with 2-plane, 2D biplane Simpson method and the reference values. Furthermore, the measurements of RV volume and free wall mass with 16-plane and 8-plane were better correlated with the reference values than those with 4-plane and 2D length method.
CONCLUSIONS: RT-3DE will be a valuable technique for quantifying irregular crescentic RV volume and wall mass.

Entities:  

Mesh:

Year:  2006        PMID: 16686622     DOI: 10.1111/j.1540-8175.2006.00221.x

Source DB:  PubMed          Journal:  Echocardiography        ISSN: 0742-2822            Impact factor:   1.724


  4 in total

1.  Manual correction of semi-automatic three-dimensional echocardiography is needed for right ventricular assessment in adults; validation with cardiac magnetic resonance.

Authors:  Ellen Ostenfeld; Marcus Carlsson; Kambiz Shahgaldi; Anders Roijer; Johan Holm
Journal:  Cardiovasc Ultrasound       Date:  2012-01-06       Impact factor: 2.062

Review 2.  Right Ventricular Dimensions and Function: Why do we Need a More Accurate and Quantitative Imaging?

Authors:  Paola Gripari; Manuela Muratori; Laura Fusini; Gloria Tamborini; Sarah Ghulam Ali; Denise Brusoni; Mauro Pepi
Journal:  J Cardiovasc Echogr       Date:  2015 Jan-Mar

3.  Quantification of right ventricular volume in dogs: a comparative study between three-dimensional echocardiography and computed tomography with the reference method magnetic resonance imaging.

Authors:  Anne K Sieslack; Peter Dziallas; Ingo Nolte; Patrick Wefstaedt; Stephan O Hungerbühler
Journal:  BMC Vet Res       Date:  2014-10-12       Impact factor: 2.741

4.  Validation of 3D echocardiographic volume detection of left atrium by human cadaveric casts.

Authors:  Jouni K Kuusisto; Vesa M Järvinen; Juha P Sinisalo
Journal:  BMC Med Imaging       Date:  2018-11-15       Impact factor: 1.930

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.