Literature DB >> 16685888

A complete augmented reality guidance system for liver punctures: first clinical evaluation.

S A Nicolau1, X Pennec, L Soler, N Ayache.   

Abstract

We provided in an augmented reality guidance system for liver punctures, which has been validated on a static abdominal phantom. In this paper, we report the first in vivo experiments. We developed a strictly passive protocol to directly evaluate our system on patients. We show that the system algorithms work efficiently and we highlight the clinical constraints that we had to overcome (small operative field, weight and sterility of the tracked marker attached to the needle...). Finally, we investigate to what extent breathing motion can be neglected for free breathing patient. Results show that the guiding accuracy, close to 1 cm, is sufficient for large targets only (above 3 cm of diameter) when the breathing motion is neglected. In the near future, we aim at validating our system on smaller targets using a respiratory gating technique.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16685888     DOI: 10.1007/11566465_67

Source DB:  PubMed          Journal:  Med Image Comput Comput Assist Interv


  1 in total

1.  Design of the image-guided biopsy marking system for gastroscopy.

Authors:  Deyu Sun; Weiling Hu; Wei Wu; Jiquan Liu; Huilong Duan; Jianmin Si
Journal:  J Med Syst       Date:  2011-08-26       Impact factor: 4.460

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.