| Literature DB >> 16681432 |
Michele Mussap1, Francesco Pietrogrande, Silvia Ponchia, Piero Maria Stefani, Roberto Sartori, Mario Plebani.
Abstract
Quantitative measurement of serum monoclonal protein (M-protein) is one of the most important tools for monitoring disease activity in monoclonal gammopathies. The aims of this study were to evaluate serum M-protein quantification by capillary zone electrophoresis (CZE) and to compare results with those obtained by densitometric scanning of high-resolution agarose gel electrophoresis (HRE-AGE). The evaluation was carried out on 82 samples from patients with various monoclonal gammopathies. All the suspected M-proteins were confirmed and characterised by immunofixation on agarose gel (IFE). CZE was performed on a Paragon CZE 2000 system (Beckman Coulter). Passing-Bablok regression was: y (CZE)=1.27 x(HRE-AGE)-5.21 g/L. The correlation coefficient was 0.92. Bland-Altman analysis demonstrated a mean difference of -1.83 g/L (95% CI -0.76 to -2.90) with clear evidence of a concentration-related bias. Densitometry gave higher values at low M-spikes (<20 g/L), whereas CZE gave higher values at large M-spikes (>20 g/L). The concentration-related bias was found to be independent of the immunoglobulin isotype. In conclusion, to compare previous results obtained by M-protein densitometric scanning with those obtained by direct measurement of CZE peaks, the calculation of a univocal transforming factor appears to be unreliable.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16681432 DOI: 10.1515/CCLM.2006.112
Source DB: PubMed Journal: Clin Chem Lab Med ISSN: 1434-6621 Impact factor: 3.694