Literature DB >> 16672023

Performance of six bone collectors designed for dental implant surgery.

Vesa T Kainulainen1, Tiina J Kainulainen, Kyösti S Oikarinen, Robert P Carmichael, George K B Sàndor.   

Abstract

OBJECTIVES: The aim of this study was to perform an in vitro comparison of six bone collectors for harvesting of particulate bone.
MATERIAL AND METHODS: Four commercially available bone collectors (Frios, Osseous Coagulum Trap, ACE Autografter, Bone Trap) and two custom-designed models were tested. Three different in vitro tests were performed to determine the harvesting capabilities of the collectors. In test I, a bovine mandible was drilled and the bone collectors were used to collect bone chips. The harvested bone volumes and dry weights were measured after harvesting. In test II, three dental implant sites were prepared in a bovine mandible. The bones from the implant osteotomies were collected, and bone volumes and dry weights were measured. In test III, 1 ml of bone chips was mixed with water, and suctioned through the bone collectors. The volumes of the bone chips retained were measured to determine the efficiency of each collector.
RESULTS: The Osseous Coagulum Trap and the custom-made collectors were the most effective instruments in test I. The mean volumes ranged from 0.17 to 0.38 ml. In test II, the difference between the collectors was small and the bone volume ranged from 0.28 to 0.37 ml. In test III, the Bone Trap became blocked before the other collectors, and its bone procurement was therefore limited.
CONCLUSION: Comparison of six different bone collectors in this in vitro study showed that all collectors are usable in clinical situations but their effectiveness varies.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16672023     DOI: 10.1111/j.1600-0501.2005.01199.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  3 in total

1.  A novel three-dimensional bone chip organ culture.

Authors:  Johannes Kuttenberger; Elzbieta Polska; Birgit M Schaefer
Journal:  Clin Oral Investig       Date:  2012-09-08       Impact factor: 3.573

2.  Application of reverse drilling technique in alveolar ridge expansion.

Authors:  Chen-Chih Chen; Ming-Dih Jeng
Journal:  J Dent Sci       Date:  2022-01-19       Impact factor: 3.719

3.  Comparison of the Bone Harvesting Capacity of an Intraoral Bone Harvesting Device and Three Different Implant Drills.

Authors:  Hyun-Chang Lim; Kyung-In Ha; Ji-Youn Hong; Ji-Young Han; Seung-Il Shin; Seung-Yun Shin; Yeek Herr; Jong-Hyuk Chung
Journal:  Biomed Res Int       Date:  2017-12-14       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.