Literature DB >> 16657177

Calibration of Infra-red CO(2) Gas Analyzers.

G C Bate1, A D'Aoust, D T Canvin.   

Abstract

Precision gas mixing pumps produce CO(2) gas mixtures for the calibration of infra-red CO(2) gas analyzers equivalent in accuracy to the standard CO(2) gas mixtures (+/- 1%) supplied by the National Bureau of Standards, Washington, D. C.The calibration of infra-red gas analyzers by the pressure difference method and by concentration differences did not agree. A factor of 0.709 was necessary to correct the pressure difference method for the effects of pressure broadening with our analyzer.

Entities:  

Year:  1969        PMID: 16657177      PMCID: PMC396226          DOI: 10.1104/pp.44.8.1122

Source DB:  PubMed          Journal:  Plant Physiol        ISSN: 0032-0889            Impact factor:   8.340


  1 in total

1.  Calibration of infra-red gas analysers for use with carbon dioxide.

Authors:  B J Legg; K J Parkinson
Journal:  J Sci Instrum       Date:  1968-10
  1 in total
  5 in total

1.  Nutrition of a developing legume fruit: functional economy in terms of carbon, nitrogen, water.

Authors:  J S Pate; P J Sharkey; C A Atkins
Journal:  Plant Physiol       Date:  1977-03       Impact factor: 8.340

2.  Comparative ecophysiology of the chromosome races in Viola adunca J.E. Smith.

Authors:  Jack Mauer; James M Mayo; Keith Denford
Journal:  Oecologia       Date:  1978-01       Impact factor: 3.225

3.  Photosynthesis and transpiration of the flag leaf in four spring-wheat cultivars.

Authors:  M Aslam; L A Hunt
Journal:  Planta       Date:  1978-01       Impact factor: 4.116

4.  Effect of pO(2) on Growth and Nodule Functioning of Symbiotic Cowpea (Vigna unguiculata L. Walp.).

Authors:  F D Dakora; C A Atkins
Journal:  Plant Physiol       Date:  1990-07       Impact factor: 8.340

5.  Effect of carbon dioxide on photorespiration.

Authors:  B A Bravdo
Journal:  Plant Physiol       Date:  1979-02       Impact factor: 8.340