Excitation of the intrinsic conduction system through his and interventricular septal pacing.

Timothy G Laske1, Nicholas D Skadsberg, Alexander J Hill, George J Klein, Paul A Iaizzo.   

Abstract

BACKGROUND: Direct His bundle pacing results in rapid synchronous ventricular activation. However, clinical experiences with such pacing have been associated with long procedure times and compromised pacing and sensing performance.
METHODS: We evaluated myocardial activation sequences (AS) for pacing of the His bundle and peri-His region and assessed acute pacing performance using custom-designed plunge electrodes. Unipolar pacing was performed in isolated swine hearts (n = 10) using four quadripolar stimulation/sensing electrodes implanted into the interventricular septum and equally spaced between the membranous septum and the coronary sinus ostium (zones 1-4, respectively; electrode depth (ED) 1 = most distal, ED 4 = most proximal). Optimal pacing sites were defined as: pacing thresholds < or = 1.5 V, a P-R ratio of < or = 0.5, and > or = 50% occurrence of an intrinsic midseptal left ventricular (LV) endocardial electrical breakout (BO) and activation pattern.
RESULTS: Pacing thresholds improved with greater depth of electrode location within the septum (ED 1: 1.51 +/- 0.8 V vs ED 4: 5.2 +/- 3.8 V, P < 0.001), as did the P-R ratio (0.34 +/- 0.6 vs 0.78 +/- 1.0, P < 0.05). His potentials were only observed in zone 1 and 2 electrodes (0.12 and 0.02 mV, respectively). Only electrodes in zones 1 and 2 produced LV endocardial electrical BOs in the midseptal region that demonstrated an intrinsic-like endocardial AS. Depth 1 and 2 electrodes (11.75 and 8.75 mm, respectively) in zone 1 satisfied all three optimal pacing site requirements.
CONCLUSIONS: This study has shown that LV activation patterns similar to sinus rhythm may be achieved without direct activation of the His bundle, while maintaining acceptable pacing and sensing performance. These data indicate that pacing systems designed to stimulate the tissues below the point at which the His bundle penetrates the central fibrous body may provide improved system efficiency and LV performance in comparison to both direct His bundle pacing and traditional pacing sites.

Entities:  

Mesh:

Year:  2006        PMID: 16650269     DOI: 10.1111/j.1540-8159.2006.00360.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  11 in total

1.  Three-dimensional imaging of ventricular activation and electrograms from intracavitary recordings.

Authors:  Chenguang Liu; Paul A Iaizzo; Bin He
Journal:  IEEE Trans Biomed Eng       Date:  2010-12-23       Impact factor: 4.538

2.  Future easy and physiological cardiac pacing.

Authors:  Eraldo Occhetta; Miriam Bortnik; Paolo Marino
Journal:  World J Cardiol       Date:  2011-01-26

Review 3.  Direct His bundle and paraHisian cardiac pacing.

Authors:  Francesco Zanon; S S Barold
Journal:  Ann Noninvasive Electrocardiol       Date:  2012-04       Impact factor: 1.468

Review 4.  [His-bundle stimulation and alternative RV stimulation sites].

Authors:  G Fröhlig; M Kindermann
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2008-03

5.  Noninvasive mapping of transmural potentials during activation in swine hearts from body surface electrocardiograms.

Authors:  Chenguang Liu; Michael D Eggen; Cory M Swingen; Paul A Iaizzo; Bin He
Journal:  IEEE Trans Med Imaging       Date:  2012-06-06       Impact factor: 10.048

6.  Right ventricular septal pacing: Safety and efficacy in a long term follow up.

Authors:  Eraldo Occhetta; Gianluca Quirino; Lara Baduena; Rosaria Nappo; Chiara Cavallino; Emanuela Facchini; Paolo Pistelli; Andrea Magnani; Miriam Bortnik; Gabriella Francalacci; Gabriele Dell'Era; Laura Plebani; Paolo Marino
Journal:  World J Cardiol       Date:  2015-08-26

7.  Estimation of global ventricular activation sequences by noninvasive three-dimensional electrical imaging: validation studies in a Swine model during pacing.

Authors:  Chenguang Liu; Nicholas D Skadsberg; Sarah E Ahlberg; Cory M Swingen; Paul A Iaizzo; Bin He
Journal:  J Cardiovasc Electrophysiol       Date:  2007-12-20

8.  Connexin 43 expression delineates two discrete pathways in the human atrioventricular junction.

Authors:  William J Hucker; Megan L McCain; Jacob I Laughner; Paul A Iaizzo; Igor R Efimov
Journal:  Anat Rec (Hoboken)       Date:  2008-02       Impact factor: 2.064

9.  Functional characteristics of left ventricular synchronization via right ventricular outflow-tract pacing detected by two-dimensional strain echocardiography.

Authors:  Yasutaka Hirayama; Yuichiro Kawamura; Nobuyuki Sato; Tatsuya Saito; Hideichi Tanaka; Yasuaki Saijo; Kenjiro Kikuchi; Katsumi Ohori; Naoyuki Hasebe
Journal:  J Arrhythm       Date:  2016-06-24

10.  Permanent parahisian pacing.

Authors:  Eraldo Occhetta; Miriam Bortnik; Paolo Marino
Journal:  Indian Pacing Electrophysiol J       Date:  2007-04-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.