C E Andersen1, C J Marckmann, M C Aznar, L Bøtter-Jensen, F Kjaer-Kristoffersen, J Medin.
Abstract
Although the radioluminescence (RL) signal from optical fibre Al(2)O(3):C dosemeters used in medical applications is essentially proportional to dose rate, the crystals used so far are imperfect in the sense that their RL sensitivity changes with accumulated dose. A computational algorithm has been developed that corrects for these sensitivity changes. We further report on a new system that effectively separates the RL signal generated in the crystal from fluorescence and Cerenkov emission generated in the optical fibre cable using a gating technique in connection with pulsed linear accelerator radiation beams. The dosimetry system has been used for dose measurements in a phantom during an intensity-modulated radiation therapy (IMRT) treatment with 6 MV photons. The RL measurement results are in excellent agreement (i.e. within 1%) with both the OSL results and the dose delivered according to the treatment planning system. RL signals from Al(2)O(3):C can be used for real-time dose rate measurements with a time resolution of approximately 0.1 s and a spatial resolution only limited by the size of the detector (<0.5 mm).Entities:
Mesh:
Substances:
Year: 2006 PMID: 16644973 DOI: 10.1093/rpd/nci600
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972