Measurements of the optical density and the thermoluminescent response of LiF:Mg,Ti exposed to high doses of 60Co gamma rays.

C Montaño-García1, I Gamboa-deBuen.   

Abstract

The absorption spectra and glow curves of LiF:Mg,Ti exposed to high doses of 60Co gamma rays as well as the optical density curves and thermoluminescent (TL) response as a function of dose were measured. Absorption peaks around 442 nm (2.8 eV), 307 nm (4.0 eV) and 248 nm (5.0 eV) were observed. The optical density is a linear function of dose up to 2500 Gy and then sublinear for the 5.0 eV band, and for the 4.0 eV band it is sublinear with a filling constant of 2.1 x 10(-4) and 6.2 x 10(-4) Gy(-1). The dose response of the 2.8 eV band is linear from 2500 Gy to 10 kGy with a slope of 1.0 x 10(-5) Gy(-1). Once the absorption spectra were obtained, the glow curves and the TL response were obtained. In the dose range, 290-10,000 Gy, the TL response of peaks 5 and 7 was supralinear-sublinear.