Literature DB >> 16644351

Analysis of factors related to successful transcatheter closure of secundum atrial septal defects using the HELEX septal occluder.

Larry A Latson1, Thomas K Jones, Joth Jacobson, Evan Zahn, John F Rhodes.   

Abstract

BACKGROUND: The HELEX septal occluder is a new transcatheter atrial septal defect occlusion device. This study analyzes the data collected from the first US clinical trials with this device to determine the most significant factors in successful use of the device.
METHODS: Tables of subsets of data were constructed in categories of possibly significant patient and device variables compared with important outcomes. Chi2 analysis was used to determine which variables had significant impact on outcomes.
RESULTS: The device was successfully implanted in 87% of patients (298/342). A major adverse event occurred in 5.8% (mostly device removal for various reasons). A significant residual leak persisted at 1 year in 2.6%. A fracture of the wire frame occurred in 8%--all were asymptomatic and only 1 required treatment. Overall composite success (no significant leak and no major adverse event) was seen in 91.5%. Patient age, body surface area, and device size had no significant influence on outcomes, except that wire frame fractures were more likely with the 35-mm device. Defect stretch diameter had the largest influence on outcomes, and implantation was possible in only 67% if the stretched diameter of the defect was >20 mm. Device-defect ratio also had a significant effect on delivery success and composite success.
CONCLUSIONS: The HELEX septal occluder is best suited for small to moderate atrial septal defects with a stretch diameter <20 mm. Although composite success was achieved in 80% of patients when a device-defect ratio <1.6 was necessitated or elected, use of a device that provided a ratio >2 resulted in 95% composite success, and is recommended when possible.

Entities:  

Mesh:

Year:  2006        PMID: 16644351     DOI: 10.1016/j.ahj.2006.01.005

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  10 in total

Review 1.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

2.  Unusual atrial septal anatomy resulting in an interatrial chamber: the true triatrial heart?

Authors:  A J Javois; D A Roberson
Journal:  Pediatr Cardiol       Date:  2007 May-Jun       Impact factor: 1.655

Review 3.  State-of-the-Art Atrial Septal Defect Closure Devices for Congenital Heart.

Authors:  Michael L O'Byrne; Daniel S Levi
Journal:  Interv Cardiol Clin       Date:  2019-01

Review 4.  The causes of Charcot-Marie-Tooth disease.

Authors:  P Young; U Suter
Journal:  Cell Mol Life Sci       Date:  2003-12       Impact factor: 9.261

5.  Helex septal occluder: feasibility study of closure of atrial septal defect.

Authors:  Amal Mahmoud El-Sisi; Salwa Gendi; Mohammed Dilawar; Mohamed Numan
Journal:  Pediatr Cardiol       Date:  2007-09-25       Impact factor: 1.655

6.  The influence of deficient retro-aortic rim on technical success and early adverse events following device closure of secundum atrial septal defects: An Analysis of the IMPACT Registry®.

Authors:  Michael L O'Byrne; Matthew J Gillespie; Kevin F Kennedy; Yoav Dori; Jonathan J Rome; Andrew C Glatz
Journal:  Catheter Cardiovasc Interv       Date:  2016-05-18       Impact factor: 2.692

Review 7.  [Catheter interventions for congenital heart disease].

Authors:  Thomas Krasemann
Journal:  Herz       Date:  2009-01-08       Impact factor: 1.443

8.  Factors related to successful transcatheter closure of atrial septal defects using the Amplatzer septal occluder.

Authors:  Namrata Rastogi; Nigel Charles Smeeton; Shakeel Ahmed Qureshi
Journal:  Pediatr Cardiol       Date:  2009-05-30       Impact factor: 1.655

Review 9.  Transcatheter device closure of atrial septal defects: more to think about than just closing the hole.

Authors:  Michael L O'Byrne; Andrew C Glatz; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2018-01       Impact factor: 2.161

10.  Prevalence of deficient retro-aortic rim and its effects on outcomes in device closure of atrial septal defects.

Authors:  Michael L O'Byrne; Andrew C Glatz; Sherzana Sunderji; Aswathi E Mathew; David J Goldberg; Yoav Dori; Jonathan J Rome; Matthew J Gillespie
Journal:  Pediatr Cardiol       Date:  2014-05-14       Impact factor: 1.655

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.