Qingyu Ma1, Dong Zhang, Xiufen Gong, Yong Ma.
Abstract
This article presents both theoretical and experimental studies on the superharmonic generation and its imaging in biological tissues. A superharmonic component is defined as a summation of the third-, fourth-, and fifth-order harmonics. A superharmonic signal is produced using an 8-mm-diam, 2.5-MHz planar piston source that is excited by eight-cycle, 2.5-MHz tone bursts. Axial and lateral field distributions of the superharmonic component and the second harmonic are first calculated based on the nonlinear KZK model and then compared with those experimentally determined at two different source pressures of 0.5 and 1 MPa. Results indicate that the amplitude of the superharmonic component can exceed that of the second harmonic, depending on the axial distance and the fundamental pressure amplitude. Also, the 3-dB beamwidth of the superharmonic component is about 23% narrower than that of the second harmonic. Additional experiments are performed in vitro using liver and fatty tissues in transmission mode and produced two-dimensional images using the fundamental, the second harmonic, and the superharmonic signals. Although the clinical applicability of this work still needs to be assessed, these results indicate that the superharmonic image quality is better than that of the other two images.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16642863 DOI: 10.1121/1.2177570
Source DB: PubMed Journal: J Acoust Soc Am ISSN: 0001-4966 Impact factor: 1.840