PURPOSE: To quantitatively and qualitatively assess perfusion with pulse-inversion (PI) ultrasonography (US) in rabbit model of acute testicular ischemia.
MATERIALS AND METHODS: Institutional animal care committee approval was obtained. After 35 rabbits underwent unilateral spermatic cord occlusion, testicular Doppler US and contrast material-enhanced PI imaging were performed. Enhancement data yielded perfusion measurements including mean value during the first 10 seconds, mean value over entire recorded replenishment curve, and curve slope during the first 5 seconds. Calculated perfusion ratios were compared with radiolabeled microsphere-derived perfusion ratios. Two readers assessed testicular perfusion as none, possible, or definite and relative perfusion as greater to the right testis than to the left, greater to the left testis than to the right, or as equal to both testes. With kappa statistics, interobserver agreement for all imaging methods was determined. Association between qualitative perfusion categories and radiolabeled microsphere-based perfusion measurements was assessed. Quantitative and qualitative determinations of relative perfusion were compared with radiolabeled microsphere-based measurements.
RESULTS: Correlations between calculated and radiolabeled microsphere-based perfusion ratios were determined (r=0.49-0.64). Interobserver agreement for presence of perfusion was excellent (kappa=0.76), and that for relative perfusion assessment was good (kappa=0.55). Neither kappa value varied significantly with imaging method. The percentage of times a testis classified as having definite perfusion had greater perfusion as measured with radiolabeled microspheres than a testis classified as having no perfusion or possible perfusion was higher with PI imaging than with Doppler US (85%-98% vs 72%-89%). Identification of the testis with less perfusion was better with quantitative methods than with qualitative assessment of images by the readers (75%-79% vs 34%-60%, P<.004).
CONCLUSION: PI imaging, compared with conventional Doppler US methods, provides superior assessment of perfusion in the setting of acute testicular ischemia. Copyright (c) RSNA, 2006.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16641337     DOI: 10.1148/radiol.2393050210

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  11 in total

1.  Assessment of tissue perfusion by contrast-enhanced ultrasound.

Authors:  Emilio Quaia
Journal:  Eur Radiol       Date:  2010-10-07       Impact factor: 5.315

Review 2.  Microbubble ultrasound contrast agents: an update.

Authors:  Emilio Quaia
Journal:  Eur Radiol       Date:  2007-03-10       Impact factor: 5.315

3.  Blood perfusion of the contralateral testis evaluated with contrast-enhanced ultrasound in rabbits with unilateral testicular torsion.

Authors:  Lin Chen; Wei-Wei Zhan; Zhou-Jun Shen; Wen-Bin Rui; Chen Lv; Man Chen; Jian-Qiao Zhou; Ping Zhou; Mi Zhou; Ying Zhu
Journal:  Asian J Androl       Date:  2009-01-19       Impact factor: 3.285

Review 4.  Safety of contrast-enhanced ultrasound in children for non-cardiac applications: a review by the Society for Pediatric Radiology (SPR) and the International Contrast Ultrasound Society (ICUS).

Authors:  Kassa Darge; Frederica Papadopoulou; Aikaterini Ntoulia; Dorothy I Bulas; Brian D Coley; Lynn A Fordham; Harriet J Paltiel; Beth McCarville; Frank M Volberg; David O Cosgrove; Barry B Goldberg; Stephanie R Wilson; Steven B Feinstein
Journal:  Pediatr Radiol       Date:  2013-07-11

5.  Role of contrast enhanced ultrasound in acute scrotal diseases.

Authors:  Massimo Valentino; Michele Bertolotto; Lorenzo Derchi; Alessandro Bertaccini; Pietro Pavlica; Giuseppe Martorana; Libero Barozzi
Journal:  Eur Radiol       Date:  2011-06-02       Impact factor: 5.315

Review 6.  Translational research in pediatric contrast-enhanced ultrasound.

Authors:  Anush Sridharan; Misun Hwang; Shelby Kutty; M Beth McCarville; Harriet J Paltiel; Maciej Piskunowicz; Sphoorti Shellikeri; Elizabeth Silvestro; George A Taylor; Ryne A Didier
Journal:  Pediatr Radiol       Date:  2021-05-15

7.  Contrast-enhanced sonography in blunt scrotal trauma().

Authors:  R Lobianco; R Regine; M De Siero; O Catalano; C Caiazzo; A Ragozzino
Journal:  J Ultrasound       Date:  2011-10-01

8.  Application of a second-generation US contrast agent in infants and children--a European questionnaire-based survey.

Authors:  Michael Riccabona
Journal:  Pediatr Radiol       Date:  2012-10-03

9.  Contrast-enhanced, real-time volumetric ultrasound imaging of tissue perfusion: preliminary results in a rabbit model of testicular torsion.

Authors:  H J Paltiel; H M Padua; P C Gargollo; G M Cannon; A I Alomari; R Yu; G T Clement
Journal:  Phys Med Biol       Date:  2011-03-14       Impact factor: 3.609

10.  Clinical and ultrasound features of segmental testicular infarction: six-year experience from a single centre.

Authors:  Praveen Bilagi; Seshadri Sriprasad; Jane L Clarke; Maria E Sellars; Gordon H Muir; Paul S Sidhu
Journal:  Eur Radiol       Date:  2007-07-05       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.